• Having set this variable apical line at the • To cancel the DR’S CHOICE apical line return to
NOTE preferred position, e.g. at the last green bar the Settings Menu, select DR’S CHOICE and tap
(Fig. 23). OFF (Fig. 25).
en en
The apical indication line shows the file
tip position inside the canal:
• blue section:
warning zone very close to the apical
region
• green to yellow sections:
apical region
• red bar:
Fig. 13 Apical section - yellow apical foramen
File movement in the Apical Zoom is accompanied
by audio signals, which serve as additional Fig. 23 DR´S CHOICE is set Fig. 25 Tap OFF
indication of the file tip position. The interval Over instrumentation
between the beeps becomes shorter the more the Once the file tip has passed the apical foramen,
file approaches the apex. the warning red dot appears beneath the apical If DR’S CHOICE is set, the DR’S CHOICE apical line
zoom image and short warning beeps will sound is additionally shown on the main screen during
When the file tip reaches the apical foramen the (Fig. 15). measurements (Fig. 24). • To exit DR’S CHOICE, tap the back arrow icon or
indication line is marked red and a constant sound the RAYPEX 6 “home” icon (Fig. 26).
®
is emitted (Fig. 14).
Fig. 15 Apical foramen is passed - red dot Fig. 24 Main screen with DR’S CHOICE set
Fig. 14 Apical foramen – red bar
When the file tip reaches the DR’S CHOICE apical
line and during further advance of the file special
WARNING beeps sound, clearly distinguished from the
As with all electronic length determination regular beeps.
devices, the bars shown in the apical When the apical foramen is reached, a solid tone Fig. 26 Exit DR’S CHOICE
zoom do not represent the distance in sounds as usual. If overinstrumentation occurs, an
millimeters. audio warning signal sounds.
Endo Easy Efficient ®