 In addition, when you use any of the other equipment   We accept no liability in the event that these   7.2 Installation
 that you have in your practice, make sure you   instructions are disregarded or processes that
 follow the directions for use for this equipment.   have not been validated are used to prepare the   Y-MK1-431  7.2.1  Charger
 en  As part of your responsibility for ensuring that the   accessories for reuse.  en
 accessories are sterile, always make sure that only   • Select the charger plug adapter which matches
 validated cleaning, disinfection and sterilization   Make sure undamaged sterile packaging is used.  your power outlet (Fig. 2).
 methods are used, that the equipment (sterilizer)
 is regularly maintained and inspected, and that the
 validated parameters are adhered to each time.
 In addition, ensure that you always comply with   8.2.1  Pre-treatment
 the applicable  statutory rules and  regulations   3    Europe            USA           Australia           UK
 with regard to hygiene in your practice or clinic. In   • Pulp and dentine remnants must be removed
 particular, this applies to the guidelines regarding   immediately from the accessories; if necessary,
 effective prion inactivation.  use a brush to carefully clean the products (no later
 For your own safety, when handling contaminated   than 2 hours after use). Do not allow remnants to   Fig. 3  Battery symbol is flashing red
 dry on. After using the accessories on the patient,
 accessories, always wear gloves, protective   place them directly into a dish filled with a suitable
 eyewear and a surgical mask.  cleaning and disinfection solution (e.g. CIDEZYME ,
 ®
 ENZOL ®  Enzymatic  Detergent  Solutions,       NOTE
 Johnson & Johnson Medical, 0.8% for between   Fig. 2  Plug adapters for charger    The device with flashing red battery
 WARNINGS  1  minute and 2  hours) for cleaning, pre-  symbol is still functional and may be
 •  The measuring cable cannot be   disinfection and interim storage. Make sure that   • Slide the plug adapter downwards into the slots   used for several treatments before the
 autoclaved.  the products are fully immersed. Then wash the   until it locks in place with a click.  device shuts down.
 •  The use of substances and materials   accessories under running sterile, deionized water
 other than those referred to above may   or in a disinfection solution at least three times
 damage the equipment and accessories.  for one minute each time, in order to remove all   WARNINGS  Charging Screen
 •  Do not employ sterilization methods that   visible traces of contamination and remnants. The   •  Prior to first use, the battery must be   Battery status during battery charging is shown on
 disinfectant should be aldehyde-free (aldehyde
 use heat, radiation, formaldehyde, ethylene   causes blood to stain permanently), tested for   charged for 6 hours!   the charging screen:
 oxide or plasma.   efficacy (e.g.  VAH/DGHM or FDA certification or    •  Use the original charger only.
 CE mark), suitable for disinfecting the accessories,   Red: charging, low battery level
 and compatible with the accessories (see section
 8.2 Cleaning, Disinfection and   8.2.7, "Material resistance").  Yellow: charging, medium battery level
 Sterilisation
 (in accordance with DIN EN ISO 17664)  7.2.2  Rechargeable Battery  Green: charging completed, battery
 • Only use clean, soft brushes to manually remove   fully charged
 contamination and remnants, or a clean, soft cloth
               ®
 The cleaning, disinfection and sterilization process   or wipe that is only used for this purpose. Do not   RAYPEX  6 is powered by a Nickel-Metal Hydride
 applies only to the lip clip, file clip and touch probe.   use  metal  brushes  or  wire  wool.  Check  that  no   (NiMH) rechargeable battery. Battery status during
 As long as they are treated with due care and are   visible contamination  or remnants remain, and   the operation is shown on the main screen:   To charge the battery, follow the next steps:
 not damaged or contaminated, these accessories   repeat the pre-cleaning process if necessary.  • Make sure that the measurement cable is not
 can be reused multiple times.   Battery Symbol  connected to the patient.
          White: indicates battery level from fully charged to   • Remove the measurement cable from RAYPEX  6.
                                                                             ®
          approximately 20% of capacity.
          Flashing red: When battery level drops under 20%,   • Connect the external charger to the device jack
                                              and plug it into the power outlet.
          the indicator will turn red and start flashing. The
          battery needs to be recharged (Fig. 3).   • If the battery is fully discharged, e.g. when it has
                                              not been used for a long period of time, it must be
                                              recharged for 6 hours.
                                              • If the battery is low, 4 hours recharging is
                                              sufficient.