Cleaning, Disinfection and Sterilization # Step-by-Step Instructions 27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Magenta 27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Cyan 27771-EYAL - 27771-EYAL.1 |
6.6.3 Apical Zone
Operation Operating mode Warning
The apical zone is divided into 3 bars graduated from 1.0 to 0.0 (apex)
• Steam sterilization at: • The accessories (lip clip,
134 °C / 273°F, hook and fork) must be – see Fig. 11.
3 min. sterilized according to the
packaging labelling.
• Use only autoclaves that
match the requirements of
EN 13060, EN 285.
• Respect the maintenance
procedure of the autoclave
device given by the
9 Sterilization
manufacturer.
• Use only this recommended
sterilization procedure.
• Control the efficiency
(packaging integrity, no
humidity, colour change of Beginning of apical zone Middle of apical zone Apical position
sterilization indicators,
physico-chemical Fig. 11
integrators, digital records of
cycles parameters). When the apex is reached, a solid tone is emitted.
• Keep devices in • Sterility cannot be The indication 0.0 on the Propex Pixi screen relates to the minor
®
sterilization guaranteed if packaging is apical foramen file position (the apical length).
packaging in a dry open, damaged or wet.
10 Storage and clean • Check the packaging and the
environment. medical devices before using 6.6.4 Over-Instrumentation
them (packaging integrity, no
humidity and validity period). A red “OVER” segment and an audio warning signal (rapid intermittent
signal) indicate that the file has passed the apex – Fig. 12.
Fig. 12