Cleaning, Disinfection and Sterilization  Cleaning, Disinfection and Sterilization



 7.2  Disinfection and Sterilization Procedure for Lip Clip,   Operation  Operating mode  Warning
 Connection Hook and Fork
                               •  Pulp and dentine   •  Do not allow remnants to dry
 Foreword                        remnants must be       on. Clean no later than 2
                                 removed from the       hours. For visible impurities
 For hygiene and sanitary safety purposes, the lip clip, the connection   accessories   that are observed on
 hook and the fork must be cleaned, disinfected and sterilized before   immediately.   instruments, a pre- cleaning
 each usage to prevent any cross-contamination between patients. This   •  After using the   is recommended by brushing
 concerns the first use as well as the subsequent uses.  accessories on the   them manually with soft
 General recommendations         patient, place them    material. Only use clean, soft
 •  Use only a disinfecting solution which is approved for its efficacy   directly into a dish   brushes to manually remove
 (VAH/DGHM-listing, CE marking, FDA approval) and in accordance   filled with a suitable   contamination and remnants,
                                 cleaning and
                                                        or a clean, soft cloth or wipe
 with the DFU of the disinfecting solution manufacturer. For all metal   disinfection solution   that is only used for this
 instruments, it is recommended to use anticorrosion disinfecting   (e.g. CIDEZYME ,   purpose. Do not use metal
                                               ®
 and cleaning agents;                  ®                brushes or wire wool. Check
 •  For your own safety, please wear personal protective equipment   ENZOL  Enzymatic   that no visible contamination
                                 Detergent Solutions,
 (gloves, glasses, mask).        Johnson & Johnson      or remnants remain, and
 •  The user is responsible for the sterility of the product for the first   Medical, 0.8% for   repeat the pre-cleaning
 cycle and each further usage as well as for the usage of damaged   2  Pre-Disinfection  between 1 minute   process if necessary.
 or dirty instruments where applicable after sterility.  and 2 hours) for   •  Make sure that the products
 •  Limitations and restrictions on reprocessing:   cleaning, pre-  •  are fully immersed.
                                                        The disinfectant solution
 the appearance of defects such as cracks, deformations (bent,   disinfection and   should be aldehyde free (to
 twisted), corrosion, loss of colour coding or marking, are indications   interim storage.  avoid blood impurities
 that the devices are not able to fulfil the intended use with the   •  Wash the   fixation), suitable for
 required safety level.          accessories under      disinfecting the accessories,
                                 flowing sterile,
 •  Use only clean water in all cleaning and rinsing steps.  deionized water or in   and compatible with the
                                 a disinfection solution   •  accessories.
                                                        Disinfectant used for pre-
 Step-by-step procedure          at least three times   treatment is only for personal
                                 for a minute each
                                 time, in order to      protection and does not
 Operation  Operating mode  Warning  remove all visible   obviate the need for
                                                        disinfection once cleaning
                                 traces of
 •  Disassemble the                                     has been completed.
 1  Disassembling                contamination and
 device.                         remnants.              Pre-treatment should never
                                                        be omitted.
                               •  Abundant rinsing    •  Use clean water.
                                 (at least 1 min).   •  If a pre-disinfectant solution
                                                        contains a corrosion
           3   Rinsing
                                                        inhibitor, it is recommended
                                                        to rinse the instruments just
                                                        before the cleaning.




 24/222  BMLPXEMDFUPRT / Rev.12 / 10-2018  A1030 000 000 00  1  BMLPXEMDFUPRT / Rev.12 / 10-2018  25/222                                                                              #                                                               27771-EYAL - 27771-EYAL.1 | 1 - A | 19-01-08 | 11:18:16 | SR:-- | Cyan   27771-EYAL -