Step-by-Step Instructions # Cleaning, Disinfection and Sterilization 1 A1030 000 000 00
6.6.5 Completion of the measurements
Operation Operating mode Warning
• Before unplugging the measurement cable from the device • Inspect devices and • Dirty instruments must be
receptacle disconnect the lip clip and the connection hook from the sort out those with cleaned and disinfected
patient. defects. again.
• Move the file stopper to the selected reference point on the tooth. • Assemble the devices • Discard instruments which
• Gently remove the file from the canal and measure the apical length (stops). show any deformations
between the stopper and the file. (bent, twisted), damages
7 Inspection (broken, corroded) or defects
Note (loss of colour coding or
As a safety precaution in order to avoid over-instrumentation, it is marking) affecting the
recommended to proceed as follows: stop file progression in the resistance, the safety or the
®
canal at the point where the Propex Pixi indicates ‘0.0’. performance of the
instrument.
Place the file onto an endodontic ruler and measure the apical length. • Do not use instrument
Subtract a minimum of 0.5 mm from the measured file length. lubricants.
• Place the devices in a • Avoid any contact between
6.7 Sound Adjustment support or container instruments during
to avoid any contact sterilization. Use supports or
®
Propex Pixi is equipped with an audio indicator which enables between instruments containers.
monitoring of the progression of the file within the canal in addition to or posts and pack the • Check the validity period of
visual monitoring. devices in the pouch given by the
8 Packaging “Sterilization manufacturer to determine
The volume can be adjusted to four different levels: mute, low, normal pouches”. the shelf life.
and high, by successively pressing the “VOLUME” button. • Use packaging which are
When sound level is muted the icon is turned off. At other sound resistant up to a temperature
levels the icon remains turned on. of 141°C (286°F) and in
accordance with
EN ISO 11607.
20/222 BMLPXEMDFUPRT / Rev.12 / 10-2018 BMLPXEMDFUPRT / Rev.12 / 10-2018 29/222