 Step-by-Step Instructions  #  Cleaning, Disinfection and Sterilization                                                                                                         1 A1030 000 000 00



 6.6.5  Completion of the measurements
                 Operation        Operating mode               Warning
 •  Before unplugging the measurement cable from the device   •  Inspect devices and   •  Dirty instruments must be
 receptacle disconnect the lip clip and the connection hook from the   sort out those with   cleaned and disinfected
 patient.                        defects.               again.
 •  Move the file stopper to the selected reference point on the tooth.  •  Assemble the devices   •  Discard instruments which
 •  Gently remove the file from the canal and measure the apical length   (stops).  show any deformations
 between the stopper and the file.                      (bent, twisted), damages
           7   Inspection                               (broken, corroded) or defects
 Note                                                   (loss of colour coding or
 As a safety precaution in order to avoid over-instrumentation, it is   marking) affecting the
 recommended to proceed as follows: stop file progression in the   resistance, the safety or the
 ®
 canal at the point where the Propex Pixi  indicates ‘0.0’.   performance of the
                                                        instrument.
 Place the file onto an endodontic ruler and measure the apical length.   •  Do not use instrument
 Subtract a minimum of 0.5 mm from the measured file length.   lubricants.
                               •  Place the devices in a   •  Avoid any contact between
 6.7  Sound Adjustment           support or container   instruments during
                                 to avoid any contact   sterilization. Use supports or
 ®
 Propex Pixi  is equipped with an audio indicator which enables   between instruments   containers.
 monitoring of the progression of the file within the canal in addition to   or posts and pack the   •  Check the validity period of
 visual monitoring.              devices in             the pouch given by the
           8   Packaging         “Sterilization         manufacturer to determine
 The volume can be adjusted to four different levels: mute, low, normal   pouches”.  the shelf life.
 and high, by successively pressing the    “VOLUME” button.   •  Use packaging which are
 When sound level is muted the    icon is turned off. At other sound   resistant up to a temperature
 levels the    icon remains turned on.                  of 141°C (286°F) and in
                                                        accordance with
                                                        EN ISO 11607.























 20/222  BMLPXEMDFUPRT / Rev.12 / 10-2018  BMLPXEMDFUPRT / Rev.12 / 10-2018  29/222