Cleaning, Disinfection and Sterilization                                                  Step-by-Step Instructions



                                                                                                 6.8     Demo Mode
               Operation       Operating mode               Warning
                            •  Once the accessories   •  The hook mechanism has to                       The built-in demo mode is available to become acquainted with the
                               have been cleaned     be activated several times                          device and to demonstrate its operation.
                               and inspected, place   during disinfection and the                        1. Disconnect the measuring cable or the charger from the device, if
                               them into the         rinse to allow the inner parts                         connected, and turn the device off.
                               disinfection bath for   to be disinfected more                            2. To start the demo mode, press and hold the    “ON / OFF” button
                               the prescribed        effectively.
                               contact time (e.g.   •  To dry use oil-free, filtered                        for about 2 seconds. until the whole display turns on (Fig. 13) and
                               Cidex OPA, Johnson    compressed air and then                                immediately turns off again accompanied by two beeps.
                               & Johnson Medical,    leave the accessories to dry
                               100% for 20 minutes);   further in a clean place for at
                               the accessories must   least 20 minutes.
                               be sufficiently    •  Once the accessories are
        5   Disinfection
                               immersed in the       dry, inspect and pack them
                               solution.             as soon as possible.
                            •  Remove the
                               accessories from the
                               disinfection bath and
                               rinse them off
                               thoroughly at least
                               five times for a minute
                               each time with water.
                            •  Dry the accessories
                               by blowing them                                                                                        Fig. 13
                               down fully.
                            •  Abundant rinsing    •  Use quality water in                               3. During the demo cycle the device operating sequence is shown on
                               (at least 1 min).     accordance with local                                  the screen.
                                                     regulations.                                        4. Demo cycles are repeated automatically until interrupted by the
                                                  •  If a disinfecting solution                             operator.
                                                     contains a corrosion                                5. To exit the demo mode press the    ”ON / OFF” button and hold for
                                                     inhibitor, it is recommended
        6   Rinsing                                                                                         about 1 second until a beep sounds. The device shuts down.
                                                     to rinse the instruments just
                                                     before the autoclaving.                             Note
                                                  •  Dry on a single use non-                                                                            ®
                                                     weaved cloth, or with a                             If the measurement cable is connected to Propex Pixi  during the
                                                     drying machine or filtered                          demo cycle, the device switches automatically to normal operation
                                                     compressed air.                                     mode.











       28/222                                        BMLPXEMDFUPRT / Rev.12 / 10-2018       #    BMLPXEMDFUPRT / Rev.12 / 10-2018                           27771-EYAL - 27771-EYAL.1 | 1 - A | 19-01-08 | 11:18:16 | SR:-- | Cyan   27771-EYAL - 27771-EYAL.1 | 1 - A | 19-01-08 | 11:18:16 | SR:-- | Yellow   27771-EYAL - 27771-EYAL.1 |