Cleaning, Disinfection and Sterilization  Cleaning, Disinfection and Sterilization



         7        Cleaning, Disinfection and Sterilization
 Operation  Operating mode  Warning
 •  Place the pre-cleaned   •  When choosing cleaning   7.1  General Recommendations
 accessories into the   agents and disinfectants,
 cleaning bath for the   make sure that:They are   •  The device does not contain user serviceable parts. The service
 prescribed contact   suitable for cleaning or   and repair should be provided by factory trained service personnel
 time (e.g.   disinfecting instruments You   only.
 ®
 CIDEZYME ,   use a disinfectant with   •  After each use, all objects that were in contact with infectious
 ®
 ENZOL  Enzymatic   proven efficacy (e.g. with   agents should be cleaned using towels impregnated with a
 Detergent Solutions,   VAH/DGHM or FDA   disinfecting and detergent solution (a bactericidal, fungicidal and
 Johnson & Johnson   certification or CE mark) and   aldehyde free solution). Use of chemical agents may cause
 Medical, 0.8% for 1   that the disinfectant is   damage to the equipment. We recommend to use only a
 minute); make sure   compatible with the cleaning   disinfecting solution which is approved for its efficacy
 that the products are   agent The chemicals used
 fully immersed (if   are compatible with the   (VAH/DGHM-listing, CE marking, FDA approval).
 necessary, use a soft   accessories.  •  The lip clip and the hook must be sterilized between treatments.
 brush to carefully   •  Combined cleaning/  Please note that the measuring cable cannot be autoclaved.
 brush them down).   disinfection products should   •  In addition, a fork is not included in the Propex Pixi  but it can be
                                                                  ®
 Manual   •  Remove the   only be used if the
 Cleaning or   accessories from the   instruments are only slightly   used and should follow the same procedure as the lip clip and hook.
 4  assisted by an   cleaning bath and   contaminated (no visible   •  Follow the “Disinfection and sterilization procedure” described in
 ultrasonic   rinse them off   contamination/remnants).   section 7.2.
 device  Comply with the   •  The user is responsible for the sterility of the lip clip, the connection
 thoroughly at least
 three times for a   concentrations and contact   hook and the fork for the first cycle and each further usage.
 minute each time with   times specified by the   •  All damaged accessories should be discarded and dirty
 sterile, deionized   manufacturers of the   accessories should be cleaned and sterilized per the procedure
 water.  cleaning agents and   described in section 7.2.
 •  Next, place the   disinfectants, as well as their
 accessories in an   instructions regarding the
 ultrasonic bath with a   intensity of subsequent
 cleaning agent (e.g.   •  rinsing.
 Use only freshly prepared
 ®
 CIDEZYME ,   solutions, water that is sterile
 ®
 ENZOL  Enzymatic   or has a low microbe content
 Detergent Solutions,   (< 10 cfu/ml) and a low
 Johnson & Johnson   endotoxin content (< 0.25
 Medical, 0.8% for 20   EU/ml, e.g. purified water
 minutes).  (PW/HPW)), and filtered, oil-
 free air for drying.