Page 23 - 27771
P. 23
Cleaning, Disinfection and Sterilization Cleaning, Disinfection and Sterilization
7 Cleaning, Disinfection and Sterilization
Operation Operating mode Warning
• Place the pre-cleaned • When choosing cleaning 7.1 General Recommendations
accessories into the agents and disinfectants,
cleaning bath for the make sure that:They are • The device does not contain user serviceable parts. The service
prescribed contact suitable for cleaning or and repair should be provided by factory trained service personnel
time (e.g. disinfecting instruments You only.
®
CIDEZYME , use a disinfectant with • After each use, all objects that were in contact with infectious
®
ENZOL Enzymatic proven efficacy (e.g. with agents should be cleaned using towels impregnated with a
Detergent Solutions, VAH/DGHM or FDA disinfecting and detergent solution (a bactericidal, fungicidal and
Johnson & Johnson certification or CE mark) and aldehyde free solution). Use of chemical agents may cause
Medical, 0.8% for 1 that the disinfectant is damage to the equipment. We recommend to use only a
minute); make sure compatible with the cleaning disinfecting solution which is approved for its efficacy
that the products are agent The chemicals used
fully immersed (if are compatible with the (VAH/DGHM-listing, CE marking, FDA approval).
necessary, use a soft accessories. • The lip clip and the hook must be sterilized between treatments.
brush to carefully • Combined cleaning/ Please note that the measuring cable cannot be autoclaved.
brush them down). disinfection products should • In addition, a fork is not included in the Propex Pixi but it can be
®
Manual • Remove the only be used if the
Cleaning or accessories from the instruments are only slightly used and should follow the same procedure as the lip clip and hook.
4 assisted by an cleaning bath and contaminated (no visible • Follow the “Disinfection and sterilization procedure” described in
ultrasonic rinse them off contamination/remnants). section 7.2.
device Comply with the • The user is responsible for the sterility of the lip clip, the connection
thoroughly at least
three times for a concentrations and contact hook and the fork for the first cycle and each further usage.
minute each time with times specified by the • All damaged accessories should be discarded and dirty
sterile, deionized manufacturers of the accessories should be cleaned and sterilized per the procedure
water. cleaning agents and described in section 7.2.
• Next, place the disinfectants, as well as their
accessories in an instructions regarding the
ultrasonic bath with a intensity of subsequent
cleaning agent (e.g. • rinsing.
Use only freshly prepared
®
CIDEZYME , solutions, water that is sterile
®
ENZOL Enzymatic or has a low microbe content
Detergent Solutions, (< 10 cfu/ml) and a low
Johnson & Johnson endotoxin content (< 0.25
Medical, 0.8% for 20 EU/ml, e.g. purified water
minutes). (PW/HPW)), and filtered, oil-
free air for drying.
26/222 BMLPXEMDFUPRT / Rev.12 / 10-2018 BMLPXEMDFUPRT / Rev.12 / 10-2018 23/222
#27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Black 27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Yellow 27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Magenta 27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Cyan #