Step-by-Step Instructions                                                                 Cleaning, Disinfection and Sterilization



       6.9     Automatic Shutdown                                                                        Operation       Operating mode               Warning
                          ®
               Propex Pixi  automatically shuts down after 3 minutes of non-use. In                                   •  Place the pre-cleaned   •  Make sure that the
               order to prolong battery life, it is recommended to switch the device off                                 accessories into the   accessories are not in direct
               after usage by pressing the    “ON / OFF” button.                                                         cleaning bath for the   contact with one another.
                                                                                                                         prescribed contact   •  The hook mechanism has to
                                                                                                                         time (e.g.            be activated during the
                                                                                                                                  ®
                                                                                                                         CIDEZYME ,            cleaning process, the rinsing
                                                                                                                               ®
                                                                                                                         ENZOL  Enzymatic      process and the ultrasonic
                                                                                                                         Detergent Solutions,   bath cleaning (press several
                                                                                                                         Johnson & Johnson     times the push button) to
                                                                                                                         Medical, 0.8% for 1   allow the inner parts to be
                                                                                                                         minute); make sure    cleaned more effectively.
                                                                                                                         that the products are   •  Make sure that the products
                                                                                                                         fully immersed (if    are fully immersed (if
                                                                                                                         necessary, use a soft   necessary, use a soft brush
                                                                                                                         brush to carefully    to carefully brush them
                                                                                                       Manual            brush them down).     down).
                                                                                                       Cleaning or    •  Remove the         •  No visible impurities should
                                                                                                   4   assisted by an    accessories from the   be observed on the
                                                                                                       ultrasonic        cleaning bath and     accessories.
                                                                                                       device            rinse them off
                                                                                                                         thoroughly at least
                                                                                                                         three times for a
                                                                                                                         minute each time with
                                                                                                                         sterile, deionized
                                                                                                                         water.
                                                                                                                      •  Next, place the
                                                                                                                         accessories in an
                                                                                                                         ultrasonic bath with a
                                                                                                                         cleaning agent (e.g.
                                                                                                                                  ®
                                                                                                                         CIDEZYME ,
                                                                                                                               ®
                                                                                                                         ENZOL  Enzymatic
                                                                                                                         Detergent Solutions,
                                                                                                                         Johnson & Johnson
                                                                                                                         Medical, 0.8% for 20
                                                                                                                         minutes).











       22/222                                        BMLPXEMDFUPRT / Rev.12 / 10-2018            BMLPXEMDFUPRT / Rev.12 / 10-2018                                  27/222





                #27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Black 27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Yellow 27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Magenta 27771-EYAL - 27771-EYAL.1 | 1 - B | 19-01-08 | 11:18:16 | SR:-- | Cyan   #