       Step-by-Step Instructions                                                 #               Cleaning, Disinfection and Sterilization                                                                                                                              1 A1030 000 000 00



       6.6.5   Completion of the measurements
                                                                                                         Operation       Operating mode               Warning
               •  Before unplugging the measurement cable from the device                                             •  Inspect devices and   •  Dirty instruments must be
                  receptacle disconnect the lip clip and the connection hook from the                                    sort out those with   cleaned and disinfected
                  patient.                                                                                               defects.              again.
               •  Move the file stopper to the selected reference point on the tooth.                                 •  Assemble the devices   •  Discard instruments which
               •  Gently remove the file from the canal and measure the apical length                                    (stops).              show any deformations
                  between the stopper and the file.                                                                                            (bent, twisted), damages
                                                                                                   7   Inspection                              (broken, corroded) or defects
               Note                                                                                                                            (loss of colour coding or
               As a safety precaution in order to avoid over-instrumentation, it is                                                            marking) affecting the
               recommended to proceed as follows: stop file progression in the                                                                 resistance, the safety or the
                                                   ®
               canal at the point where the Propex Pixi  indicates ‘0.0’.                                                                      performance of the
                                                                                                                                               instrument.
               Place the file onto an endodontic ruler and measure the apical length.                                                       •  Do not use instrument
               Subtract a minimum of 0.5 mm from the measured file length.                                                                     lubricants.
                                                                                                                      •  Place the devices in a   •  Avoid any contact between
       6.7     Sound Adjustment                                                                                          support or container   instruments during
                                                                                                                         to avoid any contact   sterilization. Use supports or
                          ®
               Propex Pixi  is equipped with an audio indicator which enables                                            between instruments   containers.
               monitoring of the progression of the file within the canal in addition to                                 or posts and pack the   •  Check the validity period of
               visual monitoring.                                                                                        devices in            the pouch given by the
                                                                                                   8   Packaging         “Sterilization        manufacturer to determine
               The volume can be adjusted to four different levels: mute, low, normal                                    pouches”.             the shelf life.
               and high, by successively pressing the    “VOLUME” button.                                                                   •  Use packaging which are
               When sound level is muted the    icon is turned off. At other sound                                                             resistant up to a temperature
               levels the    icon remains turned on.                                                                                           of 141°C (286°F) and in
                                                                                                                                               accordance with
                                                                                                                                               EN ISO 11607.























       20/222                                        BMLPXEMDFUPRT / Rev.12 / 10-2018            BMLPXEMDFUPRT / Rev.12 / 10-2018                                  29/222