Cleaning, Disinfection and Sterilization  Step-by-Step Instructions



         6.8      Demo Mode
 Operation  Operating mode  Warning
 •  Once the accessories   •  The hook mechanism has to   The built-in demo mode is available to become acquainted with the
 have been cleaned   be activated several times   device and to demonstrate its operation.
 and inspected, place   during disinfection and the   1. Disconnect the measuring cable or the charger from the device, if
 them into the   rinse to allow the inner parts   connected, and turn the device off.
 disinfection bath for   to be disinfected more   2. To start the demo mode, press and hold the    “ON / OFF” button
 the prescribed   effectively.
 contact time (e.g.   •  To dry use oil-free, filtered   for about 2 seconds. until the whole display turns on (Fig. 13) and
 Cidex OPA, Johnson   compressed air and then   immediately turns off again accompanied by two beeps.
 & Johnson Medical,   leave the accessories to dry
 100% for 20 minutes);   further in a clean place for at
 the accessories must   least 20 minutes.
 be sufficiently   •  Once the accessories are
 5  Disinfection
 immersed in the   dry, inspect and pack them
 solution.  as soon as possible.
 •  Remove the
 accessories from the
 disinfection bath and
 rinse them off
 thoroughly at least
 five times for a minute
 each time with water.
 •  Dry the accessories
 by blowing them                              Fig. 13
 down fully.
 •  Abundant rinsing    •  Use quality water in   3. During the demo cycle the device operating sequence is shown on
 (at least 1 min).  accordance with local   the screen.
 regulations.     4. Demo cycles are repeated automatically until interrupted by the
 •  If a disinfecting solution   operator.
 contains a corrosion   5. To exit the demo mode press the    ”ON / OFF” button and hold for
 inhibitor, it is recommended
 6  Rinsing          about 1 second until a beep sounds. The device shuts down.
 to rinse the instruments just
 before the autoclaving.  Note
 •  Dry on a single use non-                                      ®
 weaved cloth, or with a   If the measurement cable is connected to Propex Pixi  during the
 drying machine or filtered   demo cycle, the device switches automatically to normal operation
 compressed air.  mode.











 28/222  BMLPXEMDFUPRT / Rev.12 / 10-2018  #  BMLPXEMDFUPRT / Rev.12 / 10-2018  27771-EYAL - 27771-EYAL.1 | 1 - A | 19-01-08 | 11:18:16 | SR:-- | Cyan   27771-EYAL - 27771-EYAL.1 | 1 - A | 19-01-08 | 11:18:16 | SR:-- | Yellow   27771-EYAL - 27771-EYAL.1 | 1 - A | 19-01-08 | 11:18:16 | SR:-- | Magenta   #27771-EYAL - 27771-EYAL.1 |