                                                                                                 7       Cleaning, Disinfection and Sterilization
               Operation       Operating mode               Warning
                            •  Place the pre-cleaned   •  When choosing cleaning                 7.1     General Recommendations
                               accessories into the   agents and disinfectants,
                               cleaning bath for the   make sure that:They are                           •  The device does not contain user serviceable parts. The service
                               prescribed contact    suitable for cleaning or                               and repair should be provided by factory trained service personnel
                               time (e.g.            disinfecting instruments You                           only.
                                        ®
                               CIDEZYME ,            use a disinfectant with                             •  After each use, all objects that were in contact with infectious
                                     ®
                               ENZOL  Enzymatic      proven efficacy (e.g. with                             agents should be cleaned using towels impregnated with a
                               Detergent Solutions,   VAH/DGHM or FDA                                       disinfecting and detergent solution (a bactericidal, fungicidal and
                               Johnson & Johnson     certification or CE mark) and                          aldehyde free solution). Use of chemical agents may cause
                               Medical, 0.8% for 1   that the disinfectant is                               damage to the equipment. We recommend to use only a
                               minute); make sure    compatible with the cleaning                           disinfecting solution which is approved for its efficacy
                               that the products are   agent The chemicals used
                               fully immersed (if    are compatible with the                                (VAH/DGHM-listing, CE marking, FDA approval).
                               necessary, use a soft   accessories.                                      •  The lip clip and the hook must be sterilized between treatments.
                               brush to carefully   •  Combined cleaning/                                   Please note that the measuring cable cannot be autoclaved.
                               brush them down).     disinfection products should                        •  In addition, a fork is not included in the Propex Pixi  but it can be
                                                                                                                                                          ®
            Manual          •  Remove the            only be used if the
            Cleaning or        accessories from the   instruments are only slightly                         used and should follow the same procedure as the lip clip and hook.
        4   assisted by an     cleaning bath and     contaminated (no visible                            •  Follow the “Disinfection and sterilization procedure” described in
            ultrasonic         rinse them off        contamination/remnants).                               section 7.2.
            device                                   Comply with the                                     •  The user is responsible for the sterility of the lip clip, the connection
                               thoroughly at least
                               three times for a     concentrations and contact                             hook and the fork for the first cycle and each further usage.
                               minute each time with   times specified by the                            •  All damaged accessories should be discarded and dirty
                               sterile, deionized    manufacturers of the                                   accessories should be cleaned and sterilized per the procedure
                               water.                cleaning agents and                                    described in section 7.2.
                            •  Next, place the       disinfectants, as well as their
                               accessories in an     instructions regarding the
                               ultrasonic bath with a   intensity of subsequent
                               cleaning agent (e.g.   •  rinsing.
                                                     Use only freshly prepared
                                        ®
                               CIDEZYME ,            solutions, water that is sterile
                                     ®
                               ENZOL  Enzymatic      or has a low microbe content
                               Detergent Solutions,   (< 10 cfu/ml) and a low
                               Johnson & Johnson     endotoxin content (< 0.25
                               Medical, 0.8% for 20   EU/ml, e.g. purified water
                               minutes).             (PW/HPW)), and filtered, oil-
                                                     free air for drying.










       26/222                                        BMLPXEMDFUPRT / Rev.12 / 10-2018            BMLPXEMDFUPRT / Rev.12 / 10-2018                                  23/222