Page 60 - Veterinary Laser Therapy in Small Animal Practice
P. 60

46                             Veterinary Laser Therapy in Small Animal Practice


                         100%                                           simulation data on a perfect slab of tissue with a per-
                                                          g=0.95        fectly cylindrical, homogeneous beam of photons inci-
                                                          g=0.6
                         10%                              g=0.3         dent perfectly perpendicular to the surface. And even in
                                                                        that case you’d see some serious spreading of the beam
                       Probability  1%                                  and decaying intensity over the first few centimeters of

                                                                        depth. In fact, Figure 6.4 is based on some of this data.
                                                                                                                   [200]
                                                                          But does that description of perfect geometry and
                         0.1%
                                                                        homogeneous tissue sound like your Beagle’s hip? Not
                                                                        even close. Also, do you plan to keep the hand-piece in
                        0.01%                                           exactly the same spot for the entire treatment? If you’re
                            -180°    -90°      -0°     90°      180°
                                                                        not too discouraged after this let-down of an answer to
                                            Exit angle                  your main question and decide to keep reading, you’ll
                     Figure 6.2 Probability of scattering at a given angle for   learn that treatment technique is not realistic either.
                     several values of the anisotropy “g” factor. The way to read   There are several pieces of good news though.


                     this figure is to pick a g factor (say 0.95 for example), and
                     follow that line (red, in this case). What this tells you is   •  Most of the light gets inside the patient (with good
                     the probability that a photon is scattered at a given angle,   technique).
                     measured from the incident angle. So you can see that for g   •  Virtually all of that light gets absorbed somewhere
                     = 0.95, about 70% of the scatter is directly forward (angle
                     θ = 0°) and at least 99% is scattered “mostly forward” or   in the body (you don’t see massive exit beams of
                     between 45° and −45°.                                 light coming out the opposite side of the body).
                                                                        •  Dose accumulates from various treatment paths
                         1                                                 (the intersection of two treatment beams, even if
                                                                           from opposite sides of the elbow for example, pro-
                                                                           vides more dose at that intersection than along the
                        0.8                                                two individual paths).
                                                                        •  Dose accumulates at depth over treatment time
                                                                           (more detail to follow in section 6.6 on Power).
                      Anisotropy (g)                                    two things: first, that WHENEVER you are treating, it
                        0.6
                                                                          So what does this mean to you as the clinician? Well,

                        0.4
                                 Hall 2012 (cells)                      is CRITICALLY important to keep the treatment head
                                                                        PERPENDICULAR to the patient at all times. If not,
                                 Ma 2005 (skin)

                                 Firbank 1993 (skull)                   much if not most of the beam will be deflected laterally
                                 Beek 1997 (muscle)                     or back at you, rendering your therapy fairly useless.
                        0.2
                                 Jacques 2008 (mean of mouse tissue)
                                 Keijzer 1989 (intima artery)
                                                                                  He used a lot of capitals here because this
                                                                                  is really important. In practice, what this
                         0
                          300    500   700   900   1100   1300  1500
                                                                                  means is illustrated in Figure 6.5.
                                         Wavelength (nm)
                     Figure 6.3 Anisotropy factors of common tissues at a variety

                     of infrared wavelengths. Based on a range of published data.  Second, the beam profile inside the patient is not AT
                     [194–199]                                          ALL what it looks like outside the patient. It doesn’t
                                                                        matter if you have a shower-head-like treatment head
                     centimeters.” Then you say, “Really? All this compli-  or a very narrow “pencil” beam. If you are treating in

                     cated analysis and that’s all you come up with?”   one spot, you are getting light inside the body across an
                        Yup. With all this pinball action of scattering   area that is much wider (in all directions) than the spot
                     photons bouncing in all directions and incrementally   size you see on the surface of the patient.
                     being absorbed along their winding paths, that’s about   This is a big deal. When you use a scalpel, you are

                     as precise as we can be. I can show you Monte Carlo   cutting where you see the blade. Even when you apply









         REDONDO PRINT (4-COL BLEED).indd   46                                                                         08/08/2019   09:47
   55   56   57   58   59   60   61   62   63   64   65