Page 81 - scales manual v1014 web
P. 81
TOLEDO PS-60
1. Place jumper W1 in cal postion
2. Turn Unit on, let power up
3. Press and Hold UNITS key until “Setup” is displayed
4. Press UNITS key until “CAL” is displayed
5. Press ZERO to display YES or NO, Select YES and press Units
6. Empty scale and press UNITS
7. “ADD XXX”, load platter with weight and press UNITS key
8. “Done” is displayed, press UNITS to restart unit.
9. Replace W1
TORREY
TORREY PC40(NEW)
1. Turn on scale, press CAL button during Countdown
2. Display shows CODE, key in 765432 and C
3. rEF 0 press C, Press C
4. Load Press C, add full cap and press C
5. Display shows END Factor XXXX, then Turn OFF, turn off scale.
TORREY PC40L/LT & MFQ
1. Push CAL switch during countdown
2. Display goes blank, Enter 791534 and press C until you hear beep
3. Add 40 lbs and press C until you hear beep
4. Wait until you see a number appear in the ToTal display (this could take 15 seconds)
5. When number appears, remove weight and cycle power
TORREY MFQ (other)
1. During countdown, enter “C” “+” “D”
2. When display goes blank, enter 914682 Then Press “C”
3. Proceed as prompted
© 2014 ScaleResource.com Page 81
1. Place jumper W1 in cal postion
2. Turn Unit on, let power up
3. Press and Hold UNITS key until “Setup” is displayed
4. Press UNITS key until “CAL” is displayed
5. Press ZERO to display YES or NO, Select YES and press Units
6. Empty scale and press UNITS
7. “ADD XXX”, load platter with weight and press UNITS key
8. “Done” is displayed, press UNITS to restart unit.
9. Replace W1
TORREY
TORREY PC40(NEW)
1. Turn on scale, press CAL button during Countdown
2. Display shows CODE, key in 765432 and C
3. rEF 0 press C, Press C
4. Load Press C, add full cap and press C
5. Display shows END Factor XXXX, then Turn OFF, turn off scale.
TORREY PC40L/LT & MFQ
1. Push CAL switch during countdown
2. Display goes blank, Enter 791534 and press C until you hear beep
3. Add 40 lbs and press C until you hear beep
4. Wait until you see a number appear in the ToTal display (this could take 15 seconds)
5. When number appears, remove weight and cycle power
TORREY MFQ (other)
1. During countdown, enter “C” “+” “D”
2. When display goes blank, enter 914682 Then Press “C”
3. Proceed as prompted
© 2014 ScaleResource.com Page 81