Page 78 - scales manual v1014 web
P. 78
TOLEDO PANTHER

1. Open up the indicator and close SW1-1
2. Press Enter and Zero simultaneously.
3. F1 will be displayed, Press ENTER.
4. Press SELECT until F1.2 is displayed
5. This is Cal units, 1 is lb, 2 is kg.
6. Press ENTER
7. Press SELECT to go to CAL, display should read CAL 1 then press ENTER, use SELECT

to toggle to 1.
8. E SCL empty scale press enter.
9. 15 CAL delay
10. Add Ld Place test weight on scale. ENTER.
11. 000000 Enter test weight value
12. 15 CAL
13. CAL d cal done
14. F1.6 x zero cal adjust 0 is skip, 1 to store, press clear to abort.
15. F1.7 x Span cal adj.,
16. Press clear to exit.

TOLEDO SPEEDWEIGH

1. Close SW1-1
2. Press and release the PRINT and ZERO keys
3. F1 is displayed
4. Press ENTER until F1.6 is displayed
5. Change to a 1, empty scale and press ENTER
6. F1.7 is displayed, change to a 1 and press ENTER
7. Key in weight value, load platform and press ENTER
8. Press PRINT until a major program block is seen (F1)
9. Press CLEAR, press PRINT
10. Flip SW1-1

TOLEDO TIGER

1. Press Calibration switch
2. Press PLU 2 4 6 8 1 3
3. “SET”, Press FIX until “-------“
4. “Add XX”, load scale with XX pounds, Press FIX
5. “SAVE”, press FIX, “Calibration Values”, press FIX

© 2014 ScaleResource.com Page 78