Page 75 - scales manual v1014 web
P. 75
METTLER TOLEDO

TOLEDO KEY ASSIGNMENTS

Function (access program blocks) (Clears data entered)Clear

Select(scroll through choices) (back up to prior parameter)Zero

Enter not used

Escape (exit and escape parameters) not used

TOLEDO COUGAR VS and 8530

1. Place cal jumper W1 and apply power
2. “- -“ is displayed
3. Key in 1 9
4. Key in 1
5. “E SCL”, remove weight and press ENTER
6. “ADD Ld”, add test weights and press ENTER
7. Key in weight value and press ENTER
8. “E SCL”, remove weight and press ENTER
9. “Cal D”, “- -“, replace jumper W1

TOLEDO HAWK

1. Insert CAL jumper on board
2. Press PRINT and ZERO simultaneously
3. F1 is displayed, press PRINT until F1.2 is displayed
4. Using the FUNCTION key, enter a 1 and press PRINT
5. Press PRINT until “ESCALE”,
6. empty scale and press PRINT
7. “ADD Load”, add weight and press PRINT
8. Using ZERO and TARE enter weight value, press PRINT
9. Press PRINT until major block is displayed (F1)
10. Press FUNCTION until Save is displayed, press PRINT