Page 80 - scales manual v1014 web
P. 80
TOLEDO 8140 Page 80
1. Jump W2
2. Push TARE and CLEAR at the same time.
3. PRINT will enter and advance.
4. Advance to CAL C1.
5. Press TEST E Scale.
6. Press TEST
7. ADD LOAD, push PRINT
8. 000000
9. TEST will move to location, TARE changes number.
10. PRINT, then empty, PRINT.
11. CAL Done
12. S FILE
13. TEST, Cal OFF, Remove jumper to original position.
TOLEDO 8520
1. Flip Switch, Both switches down
2. Para 18, press 1
3. Press Print
4. EScale, press print
5. ADDLOAD, place weight and press Print
6. Enter test weight value, press Print
7. CalD, - -, press Print, flip switches
TOLEDO 8571
1. Remove platter and slide setup switch
2. Press TARE until F3 is displayed
3. Press PRINT
4. Press PRINT to “E Scl”, empty scale and press PRINT
5. “ADD Ld”, place weight and press PRINT
6. Press TARE to increment digit of test weight, PRINT to move flasher
And CLEAR to finalize calibration weight number
7. “Cal D”, press PRINT to F4 and press CLEAR to get out of setup
TOLEDO 8582
1. Slide setup switch and press ENTER when “SETUP” is displayed
2. Press ENTER until “E SCL”, empty scale and press ENTER
3. “Add Weight”, add weight and press ENTER
4. Enter value and press ENTER
5. “WEIGHT OK”, “EMPTY SCL”, remove weight and press ENTER
6. “CALIB DONE”, ENTER to f6.2, Y for scale 2 calibration, N to skip
7. ENTER to “TURN SW1 OFF”, flip switch and test
© 2014 ScaleResource.com
1. Jump W2
2. Push TARE and CLEAR at the same time.
3. PRINT will enter and advance.
4. Advance to CAL C1.
5. Press TEST E Scale.
6. Press TEST
7. ADD LOAD, push PRINT
8. 000000
9. TEST will move to location, TARE changes number.
10. PRINT, then empty, PRINT.
11. CAL Done
12. S FILE
13. TEST, Cal OFF, Remove jumper to original position.
TOLEDO 8520
1. Flip Switch, Both switches down
2. Para 18, press 1
3. Press Print
4. EScale, press print
5. ADDLOAD, place weight and press Print
6. Enter test weight value, press Print
7. CalD, - -, press Print, flip switches
TOLEDO 8571
1. Remove platter and slide setup switch
2. Press TARE until F3 is displayed
3. Press PRINT
4. Press PRINT to “E Scl”, empty scale and press PRINT
5. “ADD Ld”, place weight and press PRINT
6. Press TARE to increment digit of test weight, PRINT to move flasher
And CLEAR to finalize calibration weight number
7. “Cal D”, press PRINT to F4 and press CLEAR to get out of setup
TOLEDO 8582
1. Slide setup switch and press ENTER when “SETUP” is displayed
2. Press ENTER until “E SCL”, empty scale and press ENTER
3. “Add Weight”, add weight and press ENTER
4. Enter value and press ENTER
5. “WEIGHT OK”, “EMPTY SCL”, remove weight and press ENTER
6. “CALIB DONE”, ENTER to f6.2, Y for scale 2 calibration, N to skip
7. ENTER to “TURN SW1 OFF”, flip switch and test
© 2014 ScaleResource.com