TOLEDO TSM-825a

1. Place calibration jumper, “- -“
2. Key in 1 9
3. Key in 1
4. “E SCL”, remove weight and press ENTER
5. “ADD LD”, press ENTER
6. Place weight and press key in value and press ENTER
7. “E SCL”, remove weight and press ENTER
9. “- -“, replace Jumper

TOLEDO 3025/3026/3035/3045/3055/3036

1. Place W1 in cal position
2. Press and hold ZERO and press and hold PRINT key at the same time until “F0 0”
3. Press PRINT until “E SCL”
4. Remove weight and press PRINT
5. “ADD LD”, place weight and press PRINT
6. Use TARE to increment flasher, ZERO to move it, enter in weight value
7. Press PRINT
8. “F99 0”, Press PRINT, “CAL OFF”
9. Press and hold PRINT key, then press and release ZERO key then release PRINT key

TOLEDO 8510 and 3027

Calibrate as you would the previous scale, 3025 series, except you can press and hold ZERO then
press PRINT to get in and out of Cal mode

TOLEDO 8505

1. Press CAL button on right side of scale indicator
2. “10 0” is displayed, Press TARE key to change the 0 to a 1 and press PRINT
3. Press PRINT key to advance to 16
4. Press TARE key to change 0 to 1 and press PRINT
5. “E SCL” empty scale and press PRINT
6. “17 0”, change 0 to 1 by pressing TARE then press PRINT
7. “SPAN” press PRINT
8. Enter test weight value using TARE and ZERO then press PRINT
9. “Add Ld”, place weight and press PRINT
10. PRINT to 99 then press cal switch