‰  If dirty, clean exterior of feeding pump with a damp soapy cloth.
        ‰  Store feeding supplement at room temperature.  Refrigerate any opened unused liquid nutrition that is not in
            use.
        ‰  If alarm-troubleshooting guide reads, “contact service representative”, contact Norco immediately.
        ‰  Functional check schedule:  Performed every six (6) months.
        ‰  Preventive Maintenance Schedule: Performed annually.


                               HIGH FLOW HEATED HUMIDIFIED OXYGEN DELIVERY

                                                         Education
        ‰  Discuss the purpose of High Flow Heated Oxygen delivery: 1) To warm and humidify high flow oxygen
            delivery systems >8lpm.  2) To treat hypoxemia (decreased Pa02/Sa02).  3) To prevent severe nasal
            congestion/dryness associated with high flow oxygen delivery.  4) To treat hypothermia.
        ‰  Discuss contraindications of High Flow Heated Oxygen delivery:  1) Patients with occluded nostrils or
            severe septal defects should consult a physician before use.  2) Any situation in which high humidity is
            contraindicated (see AARC Clinical Practice Guidelines).

                                                       Demonstration
        ‰  Discuss the functions of the heated wire circuit humidifier:  1) Power button.  2) Alarm silence button.  3)
            Temperature setting.  4) Temperature window.  5) Alarm symbols
        ‰  Attach heated wire circuit humidifier to an IV stand.
        ‰  Demonstrate circuit assembly: 1) Attach heated wire circuit humidifier to humidifier chamber with
            appropriate temperature and heated wire cables.  2) Attach humidifier chamber spike set to the 1000ml
            sterile water port.  3) Insert oxygen tubing into the gas inlet port on the heated humidifier chamber at flows
            >8lpm.  4) Depress the power button and adjust the temperature to 32-37 degrees Fahrenheit to begin the
            therapy.
        ‰  Discuss temperature adjustment: Temperature is adjusted by turning dial on the face of the humidifier.   The
            set temperature will be displayed in the monitoring window.  *If using the Fisher Paykel MR850, the
            temperature will automatically increase to a preset level.
        ‰  Explain that the actual circuit temperature will be displayed in the monitoring window.
        ‰  Discuss acquisition of supplies and follow up.

                                          Cleaning/Maintenance/Troubleshooting
        ‰  Change the heated wire circuit, humidifier chamber and high flow cannula once a week.
        ‰  The patient must be disconnected to a standard oxygen system at the prescribed flow during the circuit
            change.
        ‰  Clean the heated wire circuit humidifier with a soapy wash cloth to clear dust or debris.
        ‰  Discuss alarm troubleshooting.  See manual on heated wire humidifier for specific alarm conditions.
        ‰  Functional check schedule:  Performed annually.
        ‰  Preventive Maintenance Schedule: None

                                                     HOSPITAL BEDS

                                                         Education
        ‰  Discuss the appropriate use of full electric, semi-electric and manual beds (e.g., when should a full, semi-
            electric or manual bed be used and why).

                                                       Demonstration
        ‰  Discuss weight restrictions for standard bed (400 lbs), and heavy-duty Bariatric beds (600 lbs).