‰  Explain the patient lead cable: 1) Black lead – place lead vertically on left side of the infant’s chest parallel
            to the nipple, 2) White lead – place lead on the right side of the chest, 3) Green lead is neutral and is not
            used. However, one can use the green lead in either the white or the black ports as a back-up lead.
        ‰  The patient belt is designed for babies > 7lbs.
        ‰  Avoid radio/electromagnetic interference (cell phones, cordless phones, microwave, etc.). Keep unit at least
            three (3) feet away from the source of interference.
        ‰  Discuss acquisition of supplies, preventive maintenance of equipment and follow up.

                                          Cleaning/Maintenance/Troubleshooting
        ‰  Explain lead rotation and care: 1) Place electrode in a clean dry area of infant’s skin. 2) Rotate electrodes
            every two to three days. (Electrodes are Hydrogel adhesive. If electrodes lose their adhesive properties, wipe
            with a damp cloth. Blow on electrodes until sticky).  3) Clean infant’s skin and place the electrodes on the
            appropriate areas.
        ‰  If infant’s skin is red, rashy, or dry/flaky, contact Norco for other lead options (patient belt).
        ‰  Wash patient belt (if applicable).
        ‰  Functional check schedule:  Performed quarterly.
        ‰  Preventive Maintenance Schedule: Performed annually.