"Oil-Less” Medical Air                                                              Medical Systems


               3.0 System Operation (continued)


               3.2 Initial Start-up

               CAUTION: Complete the prestart-up procedure before continuing with the initial start-up procedure.


                                                         WARNING:
                    To prevent electrical shock, ensure that ALL electrical power to the system is OFF, including the
                   disconnect switches and H-O-A switches on the control panel.  The facility’s supply circuit breaker
                                                   should also be locked out.
                                                                                   ®
               NOTE: Do not add oil to the compressor modules. The design of the LifeLine  compressor is totally oil-less.
               It is not necessary to fill the crankcases with oil.


                                                          WARNING:
                           Ensure that all loose articles, packing material, and tools are clear of the system.
               Set the H-O-A switches are to O (OFF).

                                                     ®
               Check all voltages supplied to the LifeLine  system to ensure they are the required value and phases needed
               by the control panel.

               Open the outlet isolation valve on each compressor module.

               Open the control line isolation valve on each compressor module.

               Close the manual drain valve on each compressor module.

               Open the receiver isolation valves on the control/dryer module.

               Close the receiver bypass valve on the control/dryer module.

               Open the control air isolation valve on the control/dryer module.

               Close the control air bypass valve on the control/dryer module.

               Close the inlet and outlet valves on both dryer/filter/regulator assemblies on the control/dryer module.

               Close the outlet isolation valve on the control/dryer module.

               Apply power to the system and turn on the disconnect switches.