Rough-In Installation
Pipeline Connection
Connection REMOTE SENSOR Pressure Switch
should be made
on top of pipe DISS (Diameter Index Safety System)
connection kits are recommended for use
Horiz. run of pipeline when attaching pressure switches to
pressure/vacuum pipeline. To utilize
these kits, a 1/4-18 NPT female thread
must be provided in piping system. All
necessary components are included in
each gas specific kit as shown in Figure
10.
To pressure switch
DISS nut
Connection LOCAL SENSOR and nipple
should be made
on top of pipe
DISS male
Horiz. run of pipeline with check
body
To pipeline
DISS Kit P/N Gas Service
6-120004-10 Oxygen
6-120004-11 Nitrous Oxide
6-120004-12 Medical Air
6-120004-13 Vacuum
6-120004-14 Nitrogen
6-120004-18 Instrument Air
6-120004-19 Evac. (WAGD)
6-120004-20 Carbon Dioxide
6-120004-21 CO2/O2
6-120004-22 O2/CO2
6-120004-23 He/O2
6-120004-24 O2/He
6-120004-25 Helium
Figure 10: Pressure Switch Pipeline Connection
Figure 9: Sensor Pipeline Connection
10
Part No. 6-847684-00 Rev. E00