                   Master Alarm Power-Up
                   Chapter 1: Introduction

                                     b. The second LED (CX) indicates whether the Ethernet is connected
                                         properly.

                                           –   Solid ON indicates that the Ethernet is wired properly.
                                           –   Solid OFF indicates that the wiring is bad or that the network is
                                               not activated.
                                     c. The third LED (CY) indicates when an Ethernet packet of information
                                         is being sent or received. It flashes both when sending and receiving
                                         information.
                                           –   When the network is activated, sending and receiving
                                               information starts.

                                 5. If the Master Alarm is functioning properly, lock the front panel (DE) (see
                                     figure 1-19 on page 1-60).

                                 6. If the alarm has just been installed, or if any Area Alarm Modules
                                     communicating with the Master Alarm have been removed or replaced, do
                                     the Clear Network function (see step 7 on page 1-67.)