                                                         Configuring the Master Alarm by Using the Front Panel
                                                                                                                 1
                                                                                     Chapter 1: Introduction

                              5. When the display shows the next option,


                                 N2 Address,

                                 press the Test key to scroll from


                                 1 through 255.

                              NOTE:
                              When you get close to the number you want, scroll slowly; there is no “back”
                              button so you will have to start over.

                              NOTE:
                              A Combination Alarm that has a 30-Point Master Alarm with three Area
                              Alarm Modules cannot interface with a building automation system. The Area
                              Alarm Modules use the same connector that would otherwise connect with the
                              building automation system.

                              6. Press the Silence key to select a number.


                              The next option is the Clear Network option. Choosing Yes for this option,
                              clears the device list memory from all of the devices on the alarm network.
                              Immediately after the memory is cleared, the devices automatically re-learn the
                              device list.

                              Clear the device list when you add, remove, or replace Area Alarm Modules
                              that are communicating with the Master Alarm. This option is normally used
                              only when part of the facility is under construction.


                              CAUTION:
                              If you do not clear the alarm network after removing or exchanging
                              devices, the missing device will display as a lost communication error.

                              7. When the display shows the next option,


                                 Clear Network,

                                 press the Test key to scroll through the following choices:

                                        No
                                         Yes