Configuring the Master Alarm by Using the Front Panel
1
Chapter 1: Introduction
5. When the display shows the next option,
N2 Address,
press the Test key to scroll from
1 through 255.
NOTE:
When you get close to the number you want, scroll slowly; there is no “back”
button so you will have to start over.
NOTE:
A Combination Alarm that has a 30-Point Master Alarm with three Area
Alarm Modules cannot interface with a building automation system. The Area
Alarm Modules use the same connector that would otherwise connect with the
building automation system.
6. Press the Silence key to select a number.
The next option is the Clear Network option. Choosing Yes for this option,
clears the device list memory from all of the devices on the alarm network.
Immediately after the memory is cleared, the devices automatically re-learn the
device list.
Clear the device list when you add, remove, or replace Area Alarm Modules
that are communicating with the Master Alarm. This option is normally used
only when part of the facility is under construction.
CAUTION:
If you do not clear the alarm network after removing or exchanging
devices, the missing device will display as a lost communication error.
7. When the display shows the next option,
Clear Network,
press the Test key to scroll through the following choices:
No
Yes
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 1 - 67