Configuring the Master Alarm by Using the Front Panel
Chapter 1: Introduction
8. Press the Silence key to select one.
NOTE:
If Yes is chosen, the device list clears, and the value immediately toggles back
to No.
CAUTION:
Assign a Gas Type to all of the connected inputs. If there is an input
signal wired to the Master Alarm that has not been assigned, the
display shows: Input # - Message Not Assigned.
9. When the display shows the next option,
Input 1 - Gas Type,
press the Test key to scroll through the following choices:
Medical Air Medical Vacuum
WAGD Oxygen
Nitrogen Nitrous Oxide
Carbon Dioxide Oxy. Carb. Mix
Helium Argon
Lab Air Dental Air
Instrument Air Lab Vacuum
Dental Vacuum Unused
10. Press the Silence key to select one.
NOTE:
If Unused is selected as the Gas Type the menu goes to the next Input
number. Otherwise, the display advances to the next option, which is Message.
11. Ensure that the input signal number corresponds to the proper wiring
location as documented on the Master Alarm Input Signal Data Table (see
table 1-27 on page 1-55).
12. When the display shows the next option,
Input 1 - Message,
press the Test key to scroll through the choices (see table 1-28 on page
1-57).
Page 1 - 68 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)