                   Configuring the Master Alarm by Using the Front Panel
                   Chapter 1: Introduction
                                 8. Press the Silence key to select one.


                                 NOTE:
                                 If Yes is chosen, the device list clears, and the value immediately toggles back
                                 to No.



                                 CAUTION:
                                 Assign a Gas Type to all of the connected inputs. If there is an input
                                 signal wired to the Master Alarm that has not been assigned, the
                                 display shows: Input # - Message Not Assigned.

                                 9. When the display shows the next option,

                                     Input 1 - Gas Type,


                                     press the Test key to scroll through the following choices:
                                            Medical Air           Medical Vacuum
                                            WAGD                  Oxygen
                                            Nitrogen              Nitrous Oxide
                                            Carbon Dioxide        Oxy. Carb. Mix
                                            Helium                Argon
                                            Lab Air               Dental Air
                                            Instrument Air        Lab Vacuum
                                            Dental Vacuum         Unused


                                 10. Press the Silence key to select one.

                                 NOTE:
                                 If Unused is selected as the Gas Type the menu goes to the next Input
                                 number. Otherwise, the display advances to the next option, which is Message.

                                 11. Ensure that the input signal number corresponds to the proper wiring
                                     location as documented on the Master Alarm Input Signal Data Table (see
                                     table 1-27 on page 1-55).


                                 12. When the display shows the next option,

                                     Input 1 - Message,


                                     press the Test key to scroll through the choices (see table 1-28 on page
                                     1-57).










                   Page 1 - 68           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)