Configuring the Master Alarm by Using the Front Panel
1
Chapter 1: Introduction
13. Press the Silence key to select one.
NOTE:
If you use the Web pages to configure the Master Alarm, you can customize
messages.
When there are multiple systems for the same type of gas or vacuum,
configuring a System number helps to identify the source.
14. When the display shows the next option,
Input 1 is - System,
press the Test key to scroll through the following choices:
1 2 3
4 5 6
7 8 9
15. Press the Silence key to select one.
NOTE:
Typically, all of the alarm inputs for the same type of gas or vacuum are
mapped to the same LED. The message on the display shows the specific
source of the alarm. When you are facing the front panel, the LEDs are
numbered one through nine.
1 2 3
4 5 6
7 8 9
16. When the display shows the next option,
Input 1 lights
LED,
press the Test key to scroll through the following choices:
1 2 3
4 5 6
7 8 9
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 1 - 69