                   Configuring the Master Alarm by Using the Front Panel
                   Chapter 1: Introduction

                                         Table 1-29.  Master Alarm Front Panel Configuration Mode

                            Front Panel Item                         Description

                           Test key              Scrolls through the settings for each option.
                           Silence key           Selects the settings for each option, and then advances
                                                 to the next option. Advance more quickly by pressing
                                                 and holding.

                           Display               Shows the options and settings.



                                 NOTE:
                                 For each option, the display shows only one setting at a time.

                                 1. When the display shows the first option,

                                     Language,


                                     press the Test key to scroll through the following choices:
                                            English
                                            Francais
                                            Espanol

                                 2. Press the Silence key to select one.

                                 3. When the display shows the next option,


                                     Audible Return Time,

                                     press the Test key to scroll through the following choices:

                                            Never
                                            30 minutes
                                            60 minutes
                                            90 minutes
                                            120 minutes

                                 4. Press the Silence key to select one.


                                 NOTE:
                                 To comply with the requirements of the City of New York Department of
                                 Buildings; Materials and Equipment Acceptance Division (M.E.A.), do not set
                                 the alarm return time to “never.” Set the alarm to return every 30 minutes.

                                 The next option is for the N2 Address. This is the Ethernet network address for
                                 the optional connection to a building automation system.



                   Page 1 - 66           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)