Page 729 - Med Plaza and Cancer Center
P. 729
PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03
Local Alarm Labeling
Chapter 1: Introduction
3. Using a phillips head screwdriver, loosen the four screws (CN) enough to
slide the front panel (CO) up and off the back box (CP) (see figure 1-66 on
page 1-132).
Figure 1-66. Local Alarm Labeling
4. Slide the front panel up (CO) and off the back box (CP).
SHOCK HAZARD:
Turn off the power supply in the alarm, and leave it off during the
installation of any component or sub-assembly. Failure to do so could
result in personal injury or equipment damage.
5. Turn the power switch (CQ) to the OFF position.
6. Unplug the buzzer wiring (CR).
7. Unplug the power supply wiring (CU).
8. Remove the front panel (CO).
Page 1 - 132 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)