Page 731 - Med Plaza and Cancer Center
P. 731

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                   Local Alarm Power-Up
                   Chapter 1: Introduction


                   Local Alarm Power-Up


                                 Tools required:   Phillips head screwdriver


                                 1. Using a phillips head screwdriver, loosen the four screws (DI) enough to
                                     allow the front panel (DJ) to slide off (see figure 1-67 on page 1-134).

                                                   Figure 1-67.  Local Alarm







































                                 2. Slide the front panel (DJ) off of the back box (DL).

                                 3. Turn the power switch (DK) to the ON position.

                                 NOTE:
                                 The Local Alarm lights either a green normal LED or a red alarm LED for each
                                 of its 16 inputs. If there is an alarm condition upon power-up, the alarm
                                 sounds.

                                 4. If the Local Alarm is functioning properly, secure the front panel (DJ) to
                                     the back box (DL).








                   Page 1 - 134          MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)