PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                   Safety Tips
                   Chapter 1: Introduction



                                 WARNING:
                                 No maintenance or testing should ever be undertaken or attempted by
                                 anyone not having general experience in the repair of devices of this
                                 nature. Also, to avoid damaging the unit or any of its components, no
                                 maintenance or testing should be undertaken by qualified individuals
                                 who are not completely familiar with the procedures in this manual.
                                 Personal injury or equipment damage could occur.



                                 WARNING:
                                 Notify the facility’s personnel before downloading new software.
                                 Downloading stops the alarm’s operating mode until the process is
                                 complete. This may endanger patient welfare.



                                 WARNING:
                                 Do not remove the threaded insert from the transducer. Doing so may
                                 result in attaching the transducer to the incorrect gas or vacuum type,
                                 endangering patient welfare.



                                 WARNING:
                                 If one Master Alarm is configured to monitor different inputs from the
                                 Master Alarm that you are currently configuring, ensure that the
                                 differences are accounted for in each configuration. Failure to configure
                                 the Master Alarms correctly could endanger patient welfare.



                                 WARNING:
                                 Change the factory-programmed password after the first time you log
                                 in. Otherwise, inappropriate personnel may change the setpoints and
                                 endanger patient welfare.



                                 WARNING:
                                 Before making or breaking any gas or vacuum line connections, ensure
                                 that the system is not pressurized to avoid possible injury.