                                                                                               Safety Tips
                                                                                                                 1
                                                                                     Chapter 1: Introduction


               Safety Tips




                              WARNING:
                              Only facility-authorized personnel should troubleshoot the MedPlus
                              TotalAlert™ Alarm Network. Troubleshooting by unauthorized
                              personnel could result in personal injury or equipment damage.



                              WARNING:
                              Only facility-authorized personnel should perform preventive
                              maintenance on the MedPlus TotalAlert™ Alarm Network. Preventive
                              maintenance performed by unauthorized personnel could result in
                              personal injury or equipment damage.



                              WARNING:
                              Follow the product manufacturer’s instructions. Failure to do so could
                              result in personal injury or equipment damage.



                              WARNING:
                              After the alarm is completely installed, test for cross-connections in
                              accordance with the latest edition of the NFPA 99 standard or
                              CSA Z305.1-92 code dictated by the authority having jurisdiction.
                              Misconnection of medical gas or vacuum can result in serious or fatal
                              injury to patients.



                              WARNING:
                              In operating systems, variations in pressure can cause disruption of
                              vital life support equipment. Do not vary pressures without the
                              appropriate medical facility personnel’s permission and knowledge.
                              Varying pressure can endanger patient welfare.



                              WARNING:
                              Ensure the wiring is correct on the transformer. Failure to do so could
                              cause personal injury or equipment damage.