4.20 Remote-Mount Transducer and Demand Flow Valve Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Figure 4-21. Remote-Mount Transducer and Demand Valve Assembly
4
4. Using a phillips head screwdriver, remove the four screws (E) that secure
the lid (F).
5. Note the position of the green wire (or shield wire) (G).
NOTE:
Polarity does not need to be observed for the red and black wires.
6. Using a small screwdriver, loosen the screws (H) that secure the three
wires (G and I) to the terminal block (J).
7. Remove the wires (G and I) from the terminal block (J).
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 4 - 51