Page 854 - Med Plaza and Cancer Center
P. 854

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

               4.21  Local-Mount Transducer and DISS Demand Flow Valve Assembly for Conversion Kits (excluding
               the conversion kit housed in an Area Line Pressure Alarm (P0321) rough-in box from Hill-Rom or
               Medaes)

                         Figure 4-22. Local-Mount Transducer and DISS Demand Flow Valve
                                                                                   1
                           (three-gas area alarm rough-in box from Chemetron®  is shown)
                                              a129






























                                                                                                                 4




                              5. Loosen the screws (C) securing the three wires (A and B) to the terminal
                                 block (D) in the transducer (E).

                              6. Remove the wires (A and B) from the terminal block (D).


                              7. Disconnect the DISS nut (I) from the DISS demand flow valve (F).

                              8. Remove the transducer (E).

                              9. Disconnect and remove the DISS demand flow valve (F) from the female
                                 fitting (G) in the rough-in box (H).














                   1. Chemetron® is a registered trademark of Allied Healthcare Products, Inc.





               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)    Page 4 - 55