                   4.22  Local-Mount Transducer and DISS Demand Flow Valve Assembly for the Conversion Kit housed
                   in the Area Line Pressure Alarm (P0321) Rough-In Box from Hill-Rom or Medaes Only
                   Chapter 4: Removal, Replacement, and Adjustment Procedures
                                 5. Note the position of the green wire (or shield wire) (L) in the terminal
                                     block (M) located inside the transducer (A).

                                 NOTE:
                                 Polarity does not need to be observed for the red and black wires.

                                 6. Loosen the screws (N) securing the three wires (L and O) to the terminal
                                     block (M).

                                 7. Remove the wires (L and O) from the terminal block (M).


                                 8. Disconnect the DISS nut (P) from the DISS demand flow valve (B).

                                 9. Remove the transducer (A).

                                 10. Disconnect and remove the DISS demand flow valve (B) from the female
                                     fitting (Q) in the rough-in box (K).

                          Replacement




                                 WARNING:
                                 Ensure that the correct type of transducer and DISS demand flow valve
                                 is installed for the type of gas or vacuum being monitored. If incorrect
                                 components are installed, patient welfare may be at risk.

                                 1. Ensure that the new gas-specific transducer (A) and DISS demand flow
                                     valve (B) are the correct gas or vacuum type.

                                                                       1
                                 2. To prevent gas leaks, wrap Teflon®  tape two times around the pipe threads
                                     of the DISS demand flow valve (B).

                                 3. Thread the DISS demand flow valve (B) into the female fitting (Q) inside
                                     the rough-in box (K).



                                 CAUTION:
                                 Do not overtighten the diameter index safety system (DISS) fittings.
                                 Equipment damage or inadequate flow could occur.

                                 4. Tighten the DISS demand flow valve (B) until you feel resistance, then
                                     turn the DISS demand flow valve (B) two more full turns.




                       1. Teflon® is a registered trademark of E.I. du Pont de Nemours and Company.





                   Page 4 - 60           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)