                   4.23  Remote-Mount Transducer
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                   4.23 Remote-Mount Transducer

                                 Tools required:   Adjustable wrench
                                                   Phillips head screwdriver
                                                   Small screwdriver
                                                   Leak detection solution

                          Removal



                                 WARNING:
                                 Do not remove the threaded insert from the transducer. Doing so may
                                 result in attaching the transducer to the incorrect gas or vacuum type,
                                 endangering patient welfare.


                                 1. Leave the threaded insert (H) in the transducer (B). The transducer is not a
                                     serviceable item (see figure 4-24 on page 4-62).


                                 2. Using an adjustable wrench, remove the DISS nut (A) from the DISS
                                     demand valve (I).

                                      Figure 4-24. Remote-Mount Transducer Assembly












































                   Page 4 - 62           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)