4.23 Remote-Mount Transducer
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.23 Remote-Mount Transducer
Tools required: Adjustable wrench
Phillips head screwdriver
Small screwdriver
Leak detection solution
Removal
WARNING:
Do not remove the threaded insert from the transducer. Doing so may
result in attaching the transducer to the incorrect gas or vacuum type,
endangering patient welfare.
1. Leave the threaded insert (H) in the transducer (B). The transducer is not a
serviceable item (see figure 4-24 on page 4-62).
2. Using an adjustable wrench, remove the DISS nut (A) from the DISS
demand valve (I).
Figure 4-24. Remote-Mount Transducer Assembly
Page 4 - 62 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)