4.22 Local-Mount Transducer and DISS Demand Flow Valve Assembly for the Conversion Kit housed
in the Area Line Pressure Alarm (P0321) Rough-In Box from Hill-Rom or Medaes Only
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.22 Local-Mount Transducer and DISS Demand Flow Valve
Assembly for the Conversion Kit housed in the Area Line
Pressure Alarm (P0321) Rough-In Box from Hill-Rom or
Medaes Only
Tools required: Adjustable wrench
Phillips head screwdriver
Flat screwdriver (3/16" or smaller)
5/32" hex wrench
1
Teflon® tape
Leak detection solution
Removal
SHOCK HAZARD:
Turn off the power to the alarm, and leave it off during the installation or
replacement of any component or sub-assembly. Failure to do so could
result in personal injury or equipment damage.
1. Turn off the incoming source power to the alarm.
WARNING:
Before making or breaking any gas or vacuum line connections, ensure
that the system is not pressurized to avoid possible injury.
2. Before starting the procedure, ensure that the gas or vacuum line for the
transducer under repair is not pressurized.
3. Unlock the front panel.
NOTE:
For clarity, the front panel is not shown in the illustration for this procedure.
1. Teflon® is a registered trademark of E.I. du Pont de Nemours and Company.
Page 4 - 58 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)