                   4.20  Remote-Mount Transducer and Demand Flow Valve Assembly
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                          Replacement




                                 WARNING:
                                 The demand flow valves all fit into ¼" tee ports. Ensure that the
                                 transducer and demand flow valve assembly you install is for the type of
                                 gas or vacuum in the line you are currently connecting. If the incorrect
                                 transducer is installed, patient welfare may be compromised.

                                 1. Ensure that the new gas-specific transducer (B) and demand flow valve (C)
                                     are for the type of gas that you are currently connecting.


                                                   1
                                 2. Wrap Teflon®  tape two times around the pipe threads of the of the demand
                                     flow valve (C) threads to seal against leaks.

                                 3. Thread the demand flow valve (C) into the tee port (D).



                                 CAUTION:
                                 Do not overtighten the demand flow valve fittings. Equipment damage
                                 or inadequate flow could occur.

                                 4. Using an adjustable wrench, tighten the demand flow valve (C) until you
                                     feel resistance.


                                 5. For additional strain relief, wrap the cable around one of the screw posts
                                     (K) before making the electrical connections.

                                 6. Connect the green wire (or shield wire) (G) to the terminal block (J).


                                 7. Connect the red and black wires (I) to the terminal block (J).

                                 NOTE:
                                 Polarity does not need to be observed for the red and black wires.

                                 8. Tighten the electrical connections.

                                 9. Replace and secure the lid (F) with the screws (E).

                                 10. Connect the transducer (B) to the demand flow valve (C).


                                 11. Being careful not to damage the DISS fittings, securely tighten the demand
                                     flow valve (C).


                       1. Teflon® is a registered trademark of E.I. du Pont de Nemours and Company.





                   Page 4 - 52           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)