6.2 Setting up the Internet Interface
Chapter 6: General Procedures
Serial Port (PPP) Details
Physical Connection
Connect a computer to the Master Alarm or Area Alarm Network
Communications Board with an RS-232 cable to the proper connector.
Serial Driver
1
Windows® typically wants to dial a PPP connection. In our case, we make a
direct connection. The driver, nullmodem.inf, makes a serial connection
without dialing. Install it according to the provided instructions (see “NULL
Modem Driver Setup Instructions for Windows® Operating System” on page
6-20).
Addressing
MedPlus TotalAlert™ Alarm Network devices always use IP address
192.168.1.1 with a serial port connection. All of the Web pages that are
accessible using the Ethernet are accessible using the serial port. However, the
loading rate is much slower.
Accessing
Browsers can locate devices in the following two ways:
1. IP address
Example: http://192.168.1.1
2. Local name lookup
Example: http://Master400187
This method works on a LAN if your browser has the “Use NetBios Name
Lookup” option enabled, but will not work across routers to the Internet. The
name used is configured in the Device Setup Web page (see figure 6-10 on
page 6-18). The Device Name, which is the word Master plus its serial
number, always works.
6
1. Windows® is a registered trademark of Microsoft Corporation.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 6 - 19