                                                                          6.2  Setting up the Internet Interface
                                                                              Chapter 6: General Procedures

                       Master Alarm Module or Network Communications Board
                       Connection


                                                                                             1
                              1. Open a web browser, such as E-Microsoft Internet Explorer®  4 or
                                                       2
                                 Netscape Navigator®  4.
                              2. Check to learn if your browser uses a proxy server or not. If not, skip to
                                 step 4. If so, go to step 3.

                              3. If your Web browser uses a proxy server, it will attempt to find the network
                                 device on the Internet; so you will need to add an exception.

                                 a. To add an exception, do the following:

                                   •  While your browser is running, go to the Tools drop down menu.
                                   •  Click on the Internet Options choice.
                                   •  Click on the Connections tab.

                                   •  Click on the Lan Settings button.

                                   •  Click on the Advanced button.
                                   •  Enter 192.168.*.*, beside the HTTP: Type in the “Proxy address to
                                      use” window.
                                   •  Go to step 5.


                              4. If your Web browser does not use a proxy server, enter http://192.168.1.1/
                                 in the address window (A); then press Enter (see figure 6-2 on page 6-8).


                              NOTE:
                              This is the Controller Serial Port Address, and NEVER changes.


                              The Home page of the controller displays.

                              5. To close the session, do the following:
                                 a. Right click on the two little monitors on the task bar.

                                 b. Select Disconnect.
                                                                                                                 6









                   1. E-Microsoft Internet Explorer® is a registered trademark of Microsoft Corporation.
                   2. Netscape Navigator® is a registered trademark of Netscape Communications Corp.





               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 6 - 23