InchHg / KPA / BAR selections                                                                     InchHg



        Factory Default - InchHg

        For InchHg mode, set the switch #4 to the ON position. The LED indicating InHg located next to the
        VACUUM source reading will illuminate.

        For KPA mode, set the switch #4 to the OFF position and the switch #9 to the ON position. The LED
        indicating KPA located next to the VACUUM source reading will illuminate.                         KPA

        For BAR mode, the KPA indicating source must be changed to BAR by use of a label. Set the switch
        #4 to the OFF and the switch #9 to the OFF position. The LED indicating BAR located next to the
        VACUUM source reading will illuminate.







                                                                                                          BAR











        COMMON SETTINGS FOR PRESSURE AND VACUUM

                       Repeat Alarm Enable/Disable

                       Factory Default - Disable


                       Disable

        Set switch #5 to the OFF position to disable the repeat alarm.

        NOTE: When the repeat alarm function is disabled, the alarm will not repeat.



                       Enable


        Enable Mode: (Factory Default 30 min, when enabled).

        Set switch #5 to the ON position.

        PLEASE NOTE: The Module with the Lowest set Repeat Time is the one that controls the Repeat
        Time.  For example if one Module is set for 5min and one for 30min and both are Repeat Alarm en-
        abled, the Alarm will now Repeat every 5min.