TotalAlert Infinity™ NFPA Medical Gas Notification System
Copper tubes
1.5.2 Area Alarms connect to pressure/
vacuum pipelines
Area alarm panels include the following
modular components:
• Gas Sensor input board (B60) that can be
used for up to 8 digital sensors modules
• Digital gas sensor module (B06), can be
mounted locally in the rough-in box, re-
motely on the gas pipeline, or in a com-
patible BeaconMedæs Zone Valve Box.
Maximum of 8 digital gas sensor modules
per area alarm panel.
Sensor modules are
Area alarm panels may be configured for 1 to mounted outside alarm
8 gas inputs. If alarm is configured with gas rough-in
inputs, the sensors will be included for con- Figure 14: Remote Sensors
nection to pressure/vacuum pipeline.
Sensors may be located inside the alarm Model Number Scheme:
rough-in (local sensors) utilizing the DISS
inlet tube assemblies provided for brazing to ____ - ________ - __ - ____
the hospital pipelines.
Series - Configuration - Gas - Retrofit
Copper tubes
connect to See Model Number Chart in Section 1.6.
pressure/vacuum
pipelines Example:
Sensor modules are TotalAlert Infinity™ Area Alarm OAV
mounted inside alarm
panel box Example Model Number: T3-A10-OAV
Figure 15: Area Alarm Front Panel
Figure 13: Local Sensors
Sensors may be located outside the alarm
rough-in (remote sensors) utilizing the DISS
inlet tube assembly or by an 1/4”NPT port.
4107 9016 58.02 1-8