TotalAlert Infinity™ NFPA Medical Gas Notification System


                                                                                       Copper tubes
           1.5.2  Area Alarms                                                    connect to pressure/
                                                                                    vacuum pipelines
           Area alarm panels include the following
           modular components:

           •  Gas Sensor input board (B60) that can be
              used for up to 8 digital sensors modules
           •  Digital gas sensor module (B06), can be
              mounted locally in the rough-in box, re-
              motely on the gas pipeline, or in a com-
              patible BeaconMedæs Zone Valve Box.
              Maximum of 8 digital gas sensor modules
              per area alarm panel.
                                                                                          Sensor modules are
           Area alarm panels may be configured for 1 to                                   mounted outside alarm
           8 gas inputs. If alarm is configured with gas                                  rough-in
           inputs, the sensors will be included for con-                  Figure 14:  Remote Sensors
           nection to pressure/vacuum pipeline.
           Sensors may be located inside the alarm             Model Number Scheme:
           rough-in (local sensors) utilizing the DISS
           inlet tube assemblies provided for brazing to       ____ - ________ - __ - ____
           the hospital pipelines.
                                                               Series - Configuration - Gas - Retrofit
                                Copper tubes
                                 connect to                    See Model Number Chart in Section 1.6.
                              pressure/vacuum
                                 pipelines                     Example:
           Sensor modules are                                  TotalAlert Infinity™ Area Alarm OAV
          mounted inside alarm
               panel box                                       Example Model Number: T3-A10-OAV

























                                                                       Figure 15:  Area Alarm Front Panel
                     Figure 13:  Local Sensors

           Sensors may be located outside the alarm
           rough-in (remote sensors) utilizing the DISS
           inlet tube assembly or by an 1/4”NPT port.