TotalAlert Infinity™ NFPA Medical Gas Notification System



                                                                 •  Circuit length not to exceed 5000 feet.

                                                                 •  Cable must be twisted-pair shielded type.
                                                                    Multi-pair cables within one common
                                                                    shield are acceptable.


                                                                 2.3.3 Determining Number of
                                                                 Conductors


                                                                 The following rules along with references
                                                                 to this manual’s schematics clarify wiring
                                                                 requirements.


                   Figure 21: Zone Valve Box Location            •  Digital Gas Sensor Modules
                                                                 Two conductor cables (must be twisted pair
            2.3 Wiring                                           type with shield) are required for each Gas
                                                                 Sensor module to the Gas Input board (B60
            2.3.1 General Requirements                           or B65).


            1.  All alarms are to be powered from life           Refer to NFPA Area Wiring Diagram (Section
               safety branch of emergency power system           3.2)
               as required by applicable standards.              •  Switched signal inputs

            2.  Protect all wiring from physical damage          Two conductor cables (must be twisted pair
               by raceways or conduit as required by ap-         type with shield) are required for each signal
               plicable standards.                               between the signal input terminals and the

            3.  Wire alarm panels directly to switches or        source signal switch.
               sensors as required by applicable stan-           These signals can originate at source equip-
               dards.                                            ment or from pressure switches mounted on

            4.  Wiring runs should be made with color            main pressure / vacuum pipelines.
               coded wire.  Record color, signal, and            When two master alarms are required, the
               source of signal for each wiring lead to aid      same switch / relay contacts can be wired to
               in connection of alarm finish components.         both alarm panels. (See CAUTION below).

            5.  Avoid installing alarm panels near radio         Both pairs of wires should originate from
               transmitters, electrical motors, or switch-       switch/relay contacts.
               gear.
                                                                 Refer to NFPA Master/Combination Wiring
                                                                 Diagram (Section 3.1/3.3) for B1X, B50, B65
            2.3.2 Wire Type and Size                             signal boards.


            All low voltage wiring must meet the follow-
            ing criteria:

            •  Copper wire no smaller than 22 AWG,
               conductor insulation at least 0.010in
               (0.25mm), jacket insulation at least
               0.030in (0.76mm), rated for 300V and
               60°C (140°F) minimum.

                                                             2-3                                4107 9016 58.02