Page 7 - Zavation MIS Retractor System - Hensler Surgical 2019
P. 7

Retractor Kit                                                                                    Rev: B



                                                    USER MANUAL

                      1.3.4.  Retractor, blades and wrench


               The retractor, blades and wrench are reusable devices.

               - Choose the blades size using the graduation on the dilators (blades length depending
               on the thickness of the patient, on the  selected access for the surgery or on the

               surgeon's needs).

               - Fix the 3 selected blades on the retractor body: they slide vertically either from the top

               or from beneath, until a click can be heard:

                          Place a right blade (R marked) on the right arm (R marked) (see figure 4)

                          Place a left blade (L marked) on the left arm (L marked)
                          Place a central blade (C marked) on the middle arm

               - Before inserting the retractor  in the wound,  check that the retractor is well closed:
               blades shall make a closed circle  (see figure 4). If necessary, use the right and left

               screws to reduce the angle of the blades (refer to figure 6 for more details). Then slide
               down  the retractor body with its blades in the wound  around the largest dilator (see
               figure 5).













                            L   L                  R  R

                                         C




                                                                                  Figure 5



                                      Figure 4













                                                                                                         7