Suguna Management System Ver 1.0 / SOP / FQL / P2 - 04 Page 1 of 3
4. Estimation of Crude Protein- Kjelteck
2100 Distillation
SCOPE
To estimate the crude protein in feed &feed stuffs by
Kjelteck 2100 distillation method.
PRINCIPLE
When a solid sample in an alkaline solution is steam
distilled, some of the nitrogen in the protein will be
converted to ammonia. The ammonia will be liberated
Picture-1 Picture-2
and can be collected in a receiver solution and
estimated by titration.
Apparatus
FAnalytical balance, 1 mg or 0.1 mg accuracy
(Picture-1)
FFoss tecator 2100 Kjeltec distillation assembly
(Picture-2)
FDigestion chamber (Picture-3)
Picture-3 Picture-4
FDigestion tube (Picture-4)
FDigestion block (Picture-5)
FConical flask, 250 ml, 150 ml (Picture-6)
FVolumetric flask 1000 ml, 2000 ml (Picture-7)
FBurette, 50 ml 0.02 tolerance (Picture-8)
FHot air oven (Picture-9)
FNi-Spatula (Picture-10)
Picture-5 Picture-6
Prepared by : Approved by :
54 Director - Technical