 Troubleshooting  Cleaning, Disinfection and sterilization



         7.3      Procedure for Lip Clip, Hook and Fork
 N°  Problem  Possible cause  Solution
 5.  No sound during the   The sound control is set at   Adjust the sound level as
 procedure.  “Mute” level.  described in the User Manual.  N°  Operation  Operating mode  Warning
 Ensure a good contact between
 There is not a good contact   the mucosa and lip clip (Place   1. Do not allow remnants to dry
 between the lip clip and the oral   the lip clip in the labial angle   on. Clean no later than 2 hours.
 mucosa.                                                   For visible impurities that are
 opposite the tooth to be treated).                        observed on instruments, a pre-
 Clean the connection hook (with                           cleaning is recommended by
 The connection hook is soiled.                            brushing them manually with soft
 Ethanol).
                                    1. Pulp and dentine remnants   material.
 Indication of file position is   Deep caries provides a   must be removed from the   Only use clean, soft brushes to
 6.  not steady during the   conductive path outside the   Block the external conductive   accessories immediately.  manually remove contamination
 procedure.  canal.  path.                                 and remnants, or a clean, soft
 Remove the file, close the         2. After using the accessories on   cloth or wipe that is only used for
                                                           this purpose. Do not use metal
 perforation and repeat the apex    the patient, place them directly
 Perforation.                       into a dish filled with a suitable   brushes or wire wool. Check that
 detection procedure, carefully                            no visible contamination or
 inserting the file into canal.     cleaning and disinfection   remnants remain, and repeat the
                                                    ®
                                    solution (e.g. CIDEZYME ,
 Try continuing the procedure by                           pre-cleaning process if
                                         ®
 Large lateral canal.               ENZOL  Enzymatic Detergent
 gently advancing the file.         Solutions, Johnson & Johnson   necessary.
 Perform the cable connection   1.  Pre-Disinfection or   Medical, 0.8% for between   2. Make sure that the products
                Decontamination
 Bad electrical contact.  test as described in the User   1 minute and 2 hours) for   are fully immersed.
 Manual.                            cleaning, pre-disinfection and
                                    interim storage.       3. The disinfectant solution
 Place the connection hook on
 The connection hook is not   the metal part of the file below   should be aldehyde free (to
 properly connected to the file.    3. Wash the accessories under   avoid blood impurities fixation),
 the plastic handle.                flowing sterile, deionized water   suitable for disinfecting the
 Check the comparative X-ray        or in a disinfection solution at   accessories, and compatible
 The root canal is obliterated.     least three times for a minute
 image for hints.                                          with the accessories.
                                    each time, in order to remove all
 In the case of re-treatment: old   Remove old root filling material   visible traces of contamination   Note that disinfectant used for
 filling material residues may      and remnants.
 The device does not show   block the root canal.  residues prior to use.  pre-treatment is only for
 7.  file progression inside the                           personal protection and does not
 canal.  The root canal may be blocked   Completely remove the   obviate the need for disinfection
 by the remnants of a medication   remnants prior to use.  once cleaning has been
 (e.g. calcium hydroxide).                                 completed.
                                                           Pre-treatment should never be
 Rinse the root canal with NaCl                            omitted.
 Root canal is extremely dry.  solution. Dry the access cavity
 with a cotton pellet/ air-blower.                         Use clean water.
 If there is no parietal contact use                       If a pre-disinfectant solution
 The selected file is too small for   larger ISO size file.  2.  Rinsing  Abundant rinsing (at least   contains a corrosion inhibitor, it
 a large root canal.  Important: exactly fitting files   1 min).  is recommended to rinse the
 lead to precise results.                                  instruments just before the
                                                           cleaning.
 Electronic malfunction.  Contact your distributor.




 30/172  B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)  B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)  23/172














































































 1





 # 25270-EYAL - 25270_pnim_eyal.1.1 | 1 - A | 18-05-15 | 13:28:55 | SR:-- | Magenta  25270-EYAL - 25270_pnim_eyal.1.1 | 1 - A | 18-05-15 | 13:28:55 | SR:-- | Yellow  #25270-EYAL - 25270_pnim_eyal.1.1 | 1 - A | 18-05-15 | 13:28:55 | SR:-- | Black  25270-EYAL - 25270_pnim_eyal.1.1 | 1 - A | 18-05-15 | 13:28:55 | SR:-- | Cyan