Cleaning, Disinfection and sterilization                                        2         Technical Characteristics



                                                                                                 8       TECHNICAL CHARACTERISTICS
        N°        Operation          Operating mode            Warning
                                                        Check the validity period of the                 ProPex  II complies to IEC60601-1 safety standard and the
                                                                                                                 ®
                                                        pouch given by the manufacturer
                                                        to determine the shelf life.                     requirement of CE Marking of Conformity.
        7.    Packaging          Pack the devices in “Sterilization
                                 pouches”.              Use packaging which are
                                                        resistant up to a temperature of
                                                        141°C (286°F) and in                              0086
                                                        accordance with EN ISO 11607.
                                                                                                                 ®
                                                        The accessories (lip clip, hook                  ProPex  II electronic apex locator belongs to the following category of
                                                        and fork) must be sterilized
                                                        according to the packaging                       medical devices:
                                                        labelling.
                                                        Use fractionated vacuum or                              Specification                       Description
                                                        gravity (less preferred)
                                                        autoclaves (according to                                                           Maillefer Instruments Holding Sàrl
                                                        EN 13060, EN 285).
        8.    Sterilization      Steam sterilization at: 134°C                                                                             Chemin du Verger 3
                                 (237°F) during 3 min.  Use validated sterilization               Manufacturer                             CH-1338 Ballaigues
                                                        procedure according to                                                             Switzerland
                                                        ISO 17665-1.
                                                                                                                                           endo@dentsplysirona.com
                                                        Respect maintenance procedure
                                                                                                                                                  ®
                                                        of the autoclave device given by          Model                                    ProPex  II apex locator
                                                        the manufacturer.
                                                                                                  Dimensions (Length x Height x Width)     130 x 80 x 63 mm
                                                        Use only the listed sterilization
                                                        procedures.                               Weight                                   360 g
                                                        Sterility cannot be guaranteed if         Type of screen                           Color Graphic TFT
                                 Keep devices in sterilization
        9.    Storage            packaging in a dry and clean   packaging is open, damaged or     Screen dimensions                        3.5“
                                 environment.           wet (check the packaging before
                                                        using the instruments).
                                                                                                  Power supply                             Rechargeable battery:
                                                                                                                                           2.4 V NiMH
                                                                                                  Charger power supply (input)             120 V or 230 V
                                                                                                  Charger power supply (output)            6 VDC
                                                                                                  Frequency                                50 - 60 Hz
                                                                                                  Type of protection against electrical shock  Internally powered equipment
                                                                                                  Applied part                             BF















       26/172                            B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)  B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)            27/172














































































                                                                                       1





                                                                                                                                                                                # 25270-EYAL - 25270_pnim_eyal.1.1 | 1 - A | 18-05-15 | 13:28:55 | SR:-- | Magenta  25270-EYAL - 25270_pnim_eyal.1.1 | 1 - A | 18-05-15 | 13:28:55 | SR:-- |