Page 543 - Clinical Small Animal Internal Medicine
P. 543

49  Gastrointestinal Endoscopy  511

                 It is imperative that biopsy instruments and retrieval   water‐moistened wet gauze or toweling to remove
  VetBooks.ir  snares/baskets are held closed or within their protective   grossly contaminating material.
               outer sheath when passing through the endoscope. Open
                                                                    Complete cleaning of the endoscope requires several
               biopsy jaws or exteriorized retrieval instruments can
               puncture or otherwise damage the internal lining of the   steps and use of different cleaning products to achieve
                                                                  both physical cleaning and sterilization.
               endoscope.                                           The first step of cleaning is typically achieved with an
                 Occasionally, it is necessary to biopsy lesions that   enzymatic detergent solution. The endoscope is soaked
               require significant tip deflection for adequate visualiza-  in this solution, then the outer surfaces gently wiped
               tion. A mass on the aboral surface of the esophageal   down. The detergent solution is also pumped through
               sphincter, for instance, will require a J‐maneuver for ade-  the biopsy and air/water channels using a manifold
               quate visualization (see earlier). Attempting to pass a   (Figure 49.4) that attaches in place of the removable but-
               biopsy instrument through the endoscope while it is in   tons. If manually cleaning the endoscope, these channels
               this position is risky as the instrument is directly pushing   are  then  further  cleaned  by  passing  specific  cleaning
               against the channel lining. It is less risky to retract the   brushes.
               endoscope slightly to allow the tip to relax, pass the   After lower level cleaning with the enzymatic/deter-
               instrument until it is visualized leaving the endoscope   gent solution, higher level sterilization can be achieved
               and within the gastric lumen, then return to the maxi-  using specific sterilization fluids. These fluids are often
               mally flexed position to obtain the biopsy. Obtaining an   corrosive and toxic, and can produce toxic fumes. The
               adequate biopsy from masses in this region can be quite   operator cleaning and sterilizing the endoscope must
               difficult, as the severe turns of the endoscope restrict the   wear adequate personal protective equipment, including
               ability of the biopsy instrument to move within the   facemask and eye protection. After exposure to the steri-
               channel.                                           lization solution is complete, the endoscope is thor-
                 After a biopsy is obtained, the biopsy instrument   oughly rinsed with sterile water, both externally and
               should be kept completely closed until it has been   through the internal channels. The sterilization solutions
               entirely removed from the biopsy channel. Foreign bod-
               ies are retracted with the retrieval instrument until they
               are close to the operating end of the endoscope, then
               removed by retracting the entire endoscope.


               Endoscope Cleaning and Maintenance
               Care and maintenance of endoscopes require attention
               to detail, care with handling of the instrument, and the
               use of appropriate cleaning products and equipment.
               These are expensive and fragile pieces of equipment, and
               repair costs are high. Generally speaking, it will be best
               for only one or two specific individuals (practitioner or
               technician) to be tasked with cleaning and maintenance
               of flexible endoscopes to ensure consistency of handling
               and adequate knowledge of the necessary steps. While
               some general observations are given here, manufactur-
               er’s guidelines should be followed. This is particularly
               important with respect to cleaning fluid selection, as dif-
               ferent  manufacturers  use  differing  materials in  their
               endoscopes that may be damaged by inappropriate
               cleaning solution use.
                 Endoscopes must be thoroughly and effectively
               cleaned, on both external surfaces and within the biopsy/
               vacuum and air/water channels, after use and before
               storage. On completion of an examination, it is good
               practice to suction plain or distilled water (not saline)
               through the endoscope to clear material that may have   Figure 49.4  Attachment of a specialized manifold in place of the
                                                                  air/water and vacuum buttons allows administration of cleaning
               lodged within the working channel. The exterior body   and sterilizing solutions to the inner working channels of the
               of  the endoscope should be wiped down with        endoscope.
   538   539   540   541   542   543   544   545   546   547   548