Page 49 - Condition Monitor User Manual
P. 49

Adding Sensors




























                                      Figure 5-16 Sample Sensor Review (SSC Sensors)
                          For a BUS sensor installation (BusGuardII or BusTracII), the bus delay must be entered.
                       The bus delay is determined when the sensors are installed and must be programmed into
                       the monitor; the correct bus delay is recorded on the PD sensor termination box.  To set the
                       bus delay:
                          •  Use the control to select the correct delay in ns.

                          •  Alternately, select the ‘Delay’ link at the top of the Delay column to launch the Sensor
                              Bus Delays Assistant.  This can be used to convert the physical distance between
                              the BUS sensors to the correct delay in ns or entered in ns directly.  On pressing
                              ‘OK’, the delay values calculated are applied.






















                                           Figure 5-17 Sensor BUS Delays Assistant
                          For a PDA sensor installation (for HydroGuardII or HydroTracII):

                          •  Use the parallel control to select the correct parallel number for each sensor.
                          When an external reference is used, the reference angle between each phase and the
                       reference signal must be defined.  To set the reference angle:


                                                            43
                                                                                          www.irispower.com
   44   45   46   47   48   49   50   51   52   53   54