Page 22 - Feline diagnostic imaging
P. 22

2.3  Computed Tomography  15

                (a)                                                (b)


















               Figure 2.3  (a) Sequential scanning involves rotation of the tube followed by translation of the table. This results in acquisition of a
               dataset one slice at time. The table is advanced incrementally with each dataset acquisition. This scanning format is typically reserved
               for areas where motion will not interfere. (b) Helical scanning has simultaneous movement of the table along with gantry rotation,
               resulting in a spiral volumetric dataset. The data volume is then reconstructed into flat slices of voxels. This will allow for faster scan
               times, reducing the chance for patient motion and allowing for smaller voxel acquisition.



               a digital radiology system. It is responsible for recording   dataset. These algorithms (Figure 2.5) can be selected to
               the incident X‐ray photon and transforming that photon   display edge enhancement (bone or sharp algorithm) or for
               into an electrical signal. Modern detectors are scintillation   a  more  smooth  image  (soft  tissue  algorithm)  [2,  4]. The
               detectors in the form of crystals such as cesium iodide or   choice  of  reconstruction  algorithm  depends  on  the  area
               gadolinium oxysulfide [1, 3]. In response to X‐ray photons,   being examined. For example, a sharp or bone algorithm is
               the crystals will produce light which is electronically con-  chosen to maximize edge detection and detail when evalu-
               verted into a digital signal. The detectors are arranged over   ating osseous structures. This will result in images typically
               a  semicircular  or  annular  array,  depending  on  the  CT   more grainy. On the other hand, a smooth or soft tissue
               design. The detector configuration can be single row (sin-  algorithm  is  preferred  for  maximizing  image  contrast;
               gle slice) or multiple rows of detectors (multislice). Because   however, the images will be blurrier.
               of the increased width of the detectors in a single‐slice CT,   The attenuation data present during the CT scan is repre-
               thin  slices  require  increased  collimation,  leading  to   sented as CT number or Hounsfield units (HU) on the final
               increased scan times and tube loading [1, 3]. With multi-  image (Figure 2.6). The CT numbers are calculated based
               slice scanners, the detectors are much smaller. To allow for   on the attenuation data normalized to water [1, 2]. The CT
               variation in slice width and to decrease scan times, the sig-  number or HU for water is zero. Any structures causing
               nals from multiple rows of detector elements can be com-  more attenuation than water will have a value greater than
               bined. With multislice CT, scanning is faster and images   zero and any structures causing less attenuation will have a
               are generated with both improved temporal as well as con-  value less than zero. It is because of the range of HU values
               trast resolution. Because of the smaller slice thickness pos-  from normal and abnormal tissue that CT is superior to
               sible, spatial resolution is also improved and multiplanar   radiography for contrast resolution. Most CT scanners are
               reformatting is more accurate (Figure 2.4) [1, 2]. Although   12‐bit systems. This translates into 4096 shades of gray that
               once considered a rarity in the veterinary field, multislice   can be displayed. Each shade of gray represents a calcu-
               CT scanners have become the norm.                  lated  HU. The  typical  HU  for  various  tissues  is  listed  in
                                                                  Table 2.1. HU values range from approximately −1000 to
                                                                  +3095. Because the human eye cannot resolve this many
               2.3.1  Image Formation and Display
                                                                  shades of gray, the window and level of the gray scale are
               Computed tomography images are composed of pixels that   adjusted.  The  maximum  and  minimum  HU  values  dis-
               represent the mean attenuation value of the representative   played  are  referred  to  as  the  window  width.  The  center
               voxel. Without oversimplifying the process, the CT datasets   value of this range is referred to as the level [3]. The win-
               present  the  fraction  of  radiation  removed  while  passing   dow width and level are adjusted during the viewing pro-
               through  the  patient.  Different  mathematical  algorithms   cess  to  maximize  evaluation  of  images  (Figure  2.7).
               can be used to alter the appearance of the displayed volume   Common window width and levels for different anatomic
   17   18   19   20   21   22   23   24   25   26   27