Page 39 - Feline diagnostic imaging
P. 39

32  3  Ultrasound: Physical Principles of Ultrasound Imaging
            improved, and there is better border definition and reduced   velocities  than  pulsed  wave  Doppler  and  is  therefore
            artifact [5, 8, 9].                               needed when evaluating high‐velocity flow such as would
                                                              occur  across  with  valvular  insufficiency  jets  or  stenotic
                                                              valves during echocardiography.
            3.6   Doppler Principles                            The  Doppler  spectral  tracing  is  generated  with  both
                                                              pulsed  wave  and  continuous  wave  Doppler  techniques.
            The Doppler effect is a change in frequency caused by the   The spectral display presents flow information as a func-
            motion of a sound source, receiver, or reflector [2, 5, 6]. In   tion of time (Figure 3.6a,b) [5, 10]. Flow directed above the
            medical imaging, the Doppler effect is most often used to   baseline is toward the probe, and flow directed below the
            allow detection of the direction and velocity of blood flow.   baseline is away from the probe.
            There is a shift in frequency as sound waves are reflected by   Color  Doppler  (Figure  3.6c)  is  also  a  pulsed  Doppler
            moving targets (blood cells). If the motion is toward the   technique [5]. The Doppler shift will be color coded rather
            transducer, the frequency of the returning echoes will be   than being displayed on a spectral tracing. The clinician
            higher  than  the  transmitted  frequency.  If  the  motion  is   chooses  the  location  of  the  Doppler  evaluation  via  an
            away from the transducer, the frequency of the returning   adjustable  color  box.  The  choice  of  colors  is  operator
            echoes will be less than the transmitted frequency. This is   dependent and chosen by using a color map. One of the
            referred to as the Doppler shift. The greater the velocity, the   most commonly used color maps is known as a “BART”
            greater  the  Doppler  shift.  The  maximum  Doppler  shift   map, meaning that flow away from the transducer will be
            occurs when the sound waves are parallel to the direction   blue and flow toward the transducer will be red (Blue Away
            of flow but most vessels will not be parallel to the ultra-  Red Toward). The velocity of flow is indicated by the color
            sound beam. The Doppler angle is the angle between the   hue  or  intensity  of  the  color  [3,  5].  As  in  pulsed  wave
            direction  of  blood  flow  and  the  ultrasound  beam.  The   Doppler, the maximum velocity is limited and flow perpen-
            Doppler  shift  will  be  underestimated  if  this  angle  is  not   dicular to the ultrasound beam cannot be recorded [4, 5].
            accounted for and cannot be measured if the Doppler angle   Power  Doppler  is  sometimes  referred  to  as  energy
            is at 90°. Ideally, the Doppler angle should be between 30°   Doppler (Figure 3.6d). This is a signal processing method
            and 60° to be considered accurate [5].            that relies on the strength of the Doppler shift/signal rather
                                                              than the directional information. Power Doppler is there-
                                                              fore  dependent  on  the  amplitude  of  all  Doppler  signals
            3.7   Doppler Modes                               regardless  of  the  frequency  shift  [4,  5].  Power  Doppler
                                                              improves  the  sensitivity  to  flow  at  the  expense  of  direc-
            The most common Doppler modes utilized are pulsed wave   tional  and  quantitative  information.  Just  like  color  flow
            Doppler,  continuous  wave  Doppler,  color  Doppler,  and   Doppler, there is a color map and the intensity of the color
            power Doppler (Figure 3.6).                       is related to the signal strength. The advantage of power
              With pulsed wave Doppler (Figure 3.6a), sound is trans-  Doppler is that it is very accurate at recording very subtle or
            mitted in pulses similar to real‐time scanning [5, 6]. Pulsed   slow blood flow that could be missed with color Doppler
            wave Doppler is used in conjunction with B‐mode imaging.   evaluation [4, 5].
            Using the B‐mode image, a vessel is selected and a pulse
            gate is shown over the plane of the vessel [4]. The gate is
            the only location where the Doppler shift will be recorded.   3.8   Artifacts
            In addition to the pulse gate, a separate line superimposed
            over the gate will be visible. This line is used to estimate the   Artifacts are common in diagnostic imaging and ultrasound
            Doppler  angle  so  more  accurate  Doppler  measurements   is no exception. An imaging artifact is any misrepresenta-
            can be obtained. Pulsed wave Doppler is dependent on the   tion of the imaging anatomy. Many of these artifacts can
            PRF. As a result, flow rates are typically limited to less than   result in misleading information. A basic understanding of
            1–1.5 m/s [5].                                    these artifacts is needed if ultrasound is utilized. Some of
              Continuous wave Doppler (Figure 3.6b) utilizes a probe   the  more  common  artifacts  include  acoustic  shadowing,
            with  two  separate  sets  of  crystals  allowing  sound  to  be   distal  enhancement,  reverberation  artifact,  mirror  image
            simultaneously transmitted and received [5, 6]. Similar to   artifact, slice thickness artifact, and edge shadow artifacts.
            pulsed wave, the B‐mode image is used to select the loca-  Acoustic shadowing (Figure 3.7a) occurs when the ultra-
            tion of the continuous wave interrogation. Unlike pulsed   sound beam encounters a highly attenuating reflector such
            wave Doppler, a gate is not present and the Doppler shift   as gas or mineral that absorbs or reflects the sound [3, 5, 11].
            will be recorded across the entire line of interrogation [4].   The amplitude of the echoes beneath this reflector is reduced.
            Continuous wave Doppler can measure much higher flow   This artifact is useful in that it gives some indication as to the
   34   35   36   37   38   39   40   41   42   43   44