Page 37 - scales manual v1014 web
P. 37
ENDEL

1. Press and hold *, T, and 0 keys then turn the indicator on
2. “-“ is displayed, press the * key, “CAL” is displayed
3. Press  to enter calibration
4. Remove load and press * to display -------
5. Maximum value displayed, load scale and press *
6. Calibration is complete

FAIRBANKS SCALES

FAIRBANKS HR9000 UPS

1. Press Cal button on left side of scale under cover.
2. Press ZERO
3. Press UNITS key
4. Place weight on scale and press Cal Button

FAIRBANKS HR2200

1. Press the cal button(sw7) 3 times
2. CAL is displayed, Press the ZERO key
3. Raw counts are displayed, press the ZERO key to

establish zero
4. Press UNITS key to advance to next step
5. CAL AT will be displayed, press the ZERO key
6. Scroll in the cal weight using the UNITS (scroll) and

ZERO(enter) until calibration weight value is entered.
7. Press ZERO, LOAD is displayed
8. Load weight and press ZERO key
9. Scale Returns to gross weigh mode

© 2014 ScaleResource.com Page 37